Page 73

Teledyne API Model 465L O3 Monitor Operation Manual

5.Unplug the pump power connector from the mainboard PCA at connector J12.

6.Remove the four screws attaching the sample pump assembly to the chassis.

7.Turn over pump assembly and remove the two screws holding the sheet-metal base to the pump.

8.Install the base on the new pump.

9.Install new pump assembly in chassis.

10.Reconnect tubing, note that outlet fitting of pump should be connected to tubing routed to the ‘Exhaust’ fitting on rear panel.

11.Re-install new tubing clamps or cable-ties to secure tubing connections.

05509F DCN6409

71

Image 73
Contents Model 465L Ozone Monitor Page About Teledyne Advanced Pollution Instrumentation Tapi This page intentionally left blank General Safety Hazard Important Safety InformationMise EN Garde Consignes DE SécuritéPart Name/Description Number About this ManualRevision History Table of Contents Digital Communications Figures Sensor Detail Tables Xii 05509F DCN6409 Preface IntroductionThis page intentionally left blank Specifications Specifications and Agency ApprovalsMulti-Stream Specifications SafetyApprovals EMCPhone Other Type CertificationsThis page intentionally left blank = − InHg × ln ×  273 Principle of Operation05509F DCN6409 Getting Started Nema AC Power Connection to Monitor Pneumatic Connections Pneumatic Connections, Rack Mount ConfigurationSample Gas Connection Pneumatic Connections, Nema Configuration, Multi-StreamSample System Design in Humid Environments Sample Filtration05509F DCN6409 Permeation Gas Dryer Option Sample Conditioner OptionRear Panel Electrical I/O Connections Exhaust ConnectionElectrical I/O Connections Location of I/O Connectors in the Nema ConfigurationAnalog Output Status OutputsStatus Name Normal Operating Output # State Status Output SchematicRelay # Function Normal Operating State Hi-Current Relay OutputsRelay # Function Normal Operating StateRelay # Function Meaning Multi-Stream Relay Outputs Optional6.1. RS232 Connection Digital Serial Interfaces6.2. RS485 Connection Ethernet ConnectionHis page intentionally left blank Frequently Asked Questions This page intentionally left blank Options 05509F DCN6409 Front Panel Overview OperationOzone Concentration Display Menu Navigation KeypadAlarm Status LED Color Alarm State Concentration Alarm Status/ResetStream Selector Status/Control Optional Front Panel Menus Status LEDsMenu Navigation Basics Status LED Normal StateEditing Numeric Values Front Panel Menu Diagram OFF Analog Step-Test Diag MenuPoint 5V Output MA Output Signal Name Rear Panel Label Function Signal I/OZero Calibration Memory ResetSpan Calibration Span Cal Menu Flow Calibration Optional Electronic Flowmeter onlyFlow Cal Menu View Menu Config MenuParameter Meaning UnitsCount SETUP-VARS MenuProperty Meaning Default Range Value SETUP-STREAM MenuMulti-Stream Operation Digital Communications Changing Com Port Hardware ConfigurationDigital Communications Overview RS232 Configuration Function DB-9 General I/O Connector Pin SW2 DTE SW2 DCE RS-232 ConfigurationProperty Value Function DB-9 General I/O Connector Pin SW2 DCE RS-485 ConfigurationHardware Setup for Configuring the Ethernet Module Ethernet ConfigurationVerifying Network Hardware Connection Link LED Left Side Activity LED Right Side Color MeaningManually Assigning a Static IP Address Determining a Dynamically Assigned IP Address Dhcp ModeModbus Register Maps Modbus Commands SupportedDescription Function CodeDecimal Address Description Register Type Start Bit # Bits Real-Time Concentration and Status Registers Read OnlyInstrument Setup and Configuration Registers Read / Write Decimal Address Description Register Type Start Bit # Bits This page intentionally left blank Maintenance Schedule Maintenance and AdjustmentsMaintenance Item Recommended Interval Section Instrument Layout Instrument Layout, Rack Mount ConfigurationReplacing Internal Particulate Filter Instrument Layout, Nema ConfigurationUV Lamp Adjustment Sample Pump Replacement UV Lamp Replacement05509F DCN6409 Sensor Detail Sensor Valve ReplacementConfiguring the Analog Output Water Drop-Out Filter Option MaintenanceBoot Loader Operation Cleaning Exterior Surfaces of the M465LBoot Loader Software Setup Boot Loader Hardware SetupUpload Firmware Verify Firmware This page intentionally left blank Pneumatic Diagram TroubleshootingReference Drawings RACK-MOUNT Version only Interconnect DiagramStatus LED/Output Definitions Sensor OK Pneumatic ErrorInvalid Reading Check LampPressure Too Low Pressure Too HighThis page intentionally left blank Factory Calibration CalibrationZero Calibration This page intentionally left blank Spare Parts 12.1. M465L Rack Mount Spare Parts ListPart Number Description 12.2. M465L Nema Spare Parts ListCONNECTOR, 16 PIN, W/SCREW Flange This page intentionally left blank Permeation Gas Dryer Index