Teledyne 465L operation manual Sample Conditioner Option, Permeation Gas Dryer Option

Page 29

Teledyne API Model 465L O3 Monitor Instruction Manual

Sample Conditioner Option

The Sample Conditioner option is a water drop-out filter and a permeation gas dryer, designed for high-humidity environments where the sampling gas has been saturated with water vapor and may contain small amounts of condensed water. This coalescing filter removes not only liquid, but also particulates with a Teflon filter element. Refer to Section 9.8 for maintenance procedures.

Due to the additional volume added by the water drop-out filter, there is an exception to the Response Time in the M465L specifications:

The Response Time of an M465L instrument equipped with the Sample Conditioner option, will be <60 seconds to 95%.

CAUTION

The Sample Conditioner will NOT protect the unit from damage by large amounts of liquid, such as that ingested through immersion of the sample line.

Furthermore, it is important to closely track the amount of water accumulating in the Water Drop-out Filter after the monitor has been installed and running: if excessive condensation is occurring in the sample lines leading to the monitor, then heat-traced lines should also be used to minimize condensation

The Sample Conditioner option includes a Permeation Gas Dryer, which is available as an alternate option (described next) if humidity is the only concern and there is no risk of condensation.

Permeation Gas Dryer Option

The Permeation Gas Dryer serves to remove water vapor from sampling gas in a humid environment. It is an alternative option when there is no risk of condensation. The permeation material is a Teflon Type compound and has only a minimal effect on the ozone concentration in the sample stream. When the M465L is equipped with this option, the unit has been calibrated at the factory to compensate for any ozone loss in the permeation dryer.

05509F DCN6409

27

Image 29
Contents Model 465L Ozone Monitor Page About Teledyne Advanced Pollution Instrumentation Tapi This page intentionally left blank General Safety Hazard Important Safety InformationMise EN Garde Consignes DE SécuritéPart Name/Description Number About this ManualRevision History Table of Contents Digital Communications Figures Sensor Detail Tables Xii 05509F DCN6409 Preface IntroductionThis page intentionally left blank Specifications Specifications and Agency ApprovalsMulti-Stream Specifications SafetyApprovals EMCPhone Other Type CertificationsThis page intentionally left blank = − InHg × ln ×  273 Principle of Operation05509F DCN6409 Getting Started Nema AC Power Connection to Monitor Pneumatic Connections Pneumatic Connections, Rack Mount ConfigurationSample Gas Connection Pneumatic Connections, Nema Configuration, Multi-StreamSample System Design in Humid Environments Sample Filtration05509F DCN6409 Permeation Gas Dryer Option Sample Conditioner OptionRear Panel Electrical I/O Connections Exhaust ConnectionElectrical I/O Connections Location of I/O Connectors in the Nema ConfigurationAnalog Output Status OutputsStatus Name Normal Operating Output # State Status Output SchematicRelay # Function Normal Operating State Hi-Current Relay OutputsRelay # Function Normal Operating StateRelay # Function Meaning Multi-Stream Relay Outputs Optional6.1. RS232 Connection Digital Serial Interfaces6.2. RS485 Connection Ethernet ConnectionHis page intentionally left blank Frequently Asked Questions This page intentionally left blank Options 05509F DCN6409 Front Panel Overview OperationOzone Concentration Display Menu Navigation KeypadStream Selector Status/Control Optional Concentration Alarm Status/ResetAlarm Status LED Color Alarm State Front Panel Menus Status LEDsMenu Navigation Basics Status LED Normal StateEditing Numeric Values Front Panel Menu Diagram OFF Point 5V Output MA Output Diag MenuAnalog Step-Test Signal Name Rear Panel Label Function Signal I/OZero Calibration Memory ResetSpan Calibration Span Cal Menu Flow Calibration Optional Electronic Flowmeter onlyFlow Cal Menu View Menu Config MenuParameter Meaning UnitsCount SETUP-VARS MenuProperty Meaning Default Range Value SETUP-STREAM MenuMulti-Stream Operation Digital Communications Overview Changing Com Port Hardware ConfigurationDigital Communications RS232 Configuration Property Value RS-232 ConfigurationFunction DB-9 General I/O Connector Pin SW2 DTE SW2 DCE Function DB-9 General I/O Connector Pin SW2 DCE RS-485 ConfigurationHardware Setup for Configuring the Ethernet Module Ethernet ConfigurationVerifying Network Hardware Connection Link LED Left Side Activity LED Right Side Color MeaningManually Assigning a Static IP Address Determining a Dynamically Assigned IP Address Dhcp ModeModbus Register Maps Modbus Commands SupportedDescription Function CodeDecimal Address Description Register Type Start Bit # Bits Real-Time Concentration and Status Registers Read OnlyInstrument Setup and Configuration Registers Read / Write Decimal Address Description Register Type Start Bit # Bits This page intentionally left blank Maintenance Item Recommended Interval Section Maintenance and AdjustmentsMaintenance Schedule Instrument Layout Instrument Layout, Rack Mount ConfigurationReplacing Internal Particulate Filter Instrument Layout, Nema ConfigurationUV Lamp Adjustment Sample Pump Replacement UV Lamp Replacement05509F DCN6409 Sensor Detail Sensor Valve ReplacementConfiguring the Analog Output Water Drop-Out Filter Option MaintenanceBoot Loader Operation Cleaning Exterior Surfaces of the M465LUpload Firmware Boot Loader Hardware SetupBoot Loader Software Setup Verify Firmware This page intentionally left blank Reference Drawings TroubleshootingPneumatic Diagram RACK-MOUNT Version only Interconnect DiagramStatus LED/Output Definitions Sensor OK Pneumatic ErrorInvalid Reading Check LampPressure Too Low Pressure Too HighThis page intentionally left blank Zero Calibration CalibrationFactory Calibration This page intentionally left blank Spare Parts 12.1. M465L Rack Mount Spare Parts ListPart Number Description 12.2. M465L Nema Spare Parts ListCONNECTOR, 16 PIN, W/SCREW Flange This page intentionally left blank Permeation Gas Dryer Index