Teledyne 465L operation manual View Menu, Config Menu, Parameter Meaning, Units, Normal Range

Page 53

Teledyne API Model 465L O3 Monitor Operation Manual

7.2.4.VIEW Menu

The VIEW menu allows the operator to view various measurement parameters in real-time. This can be useful for diagnosing various instrument or system problems.

Note that these values are updated in real-time as they are repeatedly scrolled on the display.

Table 7-5: VIEW Menu Parameters

Parameter

Meaning

MEAS

UV detector reading, measure cycle

REF

UV detector reading, reference

 

cycle

PRESSURE

Sample pressure

LAMP

UV lamp temperature

TEMP

 

SAMPLE

Sample temperature

TEMP

 

FLOW

Sample Gas Flow Rate

*After initial 20 minute warm-up period.

Units

mV mV

psia ºC

ºC

cc/min

Normal Range*

250 – 1230

250 – 1230

9.0– 14.9

50.0– 52.0

5.0–50.0

640 – 960

7.2.5.CONFIG Menu

The CONFIG menu displays specific information related to the instrument’s configuration. No changes are made within this menu.

Table 7-6: CONFIG Menu Parameters

CONFIG

Meaning

S/N

Serial number of M465L. (Should

 

match the serial number tag on the rear

 

panel of the instrument).

COMM ADDRESS

Displays the MODBUS slave ID for

 

the instrument.

STREAM COUNT

Displays the number of streams

 

configured. (Always 1 for instruments

 

without the Multi-stream Option).

REV

Revision of instrument firmware

BREV

Revision of sensor firmware

BOOT

Revision of bootloader firmware

05509F DCN6409

51

Image 53
Contents Model 465L Ozone Monitor Page About Teledyne Advanced Pollution Instrumentation Tapi This page intentionally left blank General Safety Hazard Important Safety InformationMise EN Garde Consignes DE SécuritéPart Name/Description Number About this ManualRevision History Table of Contents Digital Communications Figures Sensor Detail Tables Xii 05509F DCN6409 Preface IntroductionThis page intentionally left blank Specifications Specifications and Agency ApprovalsMulti-Stream Specifications SafetyApprovals EMCPhone Other Type CertificationsThis page intentionally left blank = − InHg × ln ×  273 Principle of Operation05509F DCN6409 Getting Started Nema AC Power Connection to Monitor Pneumatic Connections Pneumatic Connections, Rack Mount ConfigurationSample Gas Connection Pneumatic Connections, Nema Configuration, Multi-StreamSample System Design in Humid Environments Sample Filtration05509F DCN6409 Permeation Gas Dryer Option Sample Conditioner OptionRear Panel Electrical I/O Connections Exhaust ConnectionElectrical I/O Connections Location of I/O Connectors in the Nema ConfigurationAnalog Output Status OutputsStatus Name Normal Operating Output # State Status Output SchematicRelay # Function Normal Operating State Hi-Current Relay OutputsRelay # Function Normal Operating StateRelay # Function Meaning Multi-Stream Relay Outputs Optional6.1. RS232 Connection Digital Serial Interfaces6.2. RS485 Connection Ethernet ConnectionHis page intentionally left blank Frequently Asked Questions This page intentionally left blank Options 05509F DCN6409 Front Panel Overview OperationOzone Concentration Display Menu Navigation KeypadStream Selector Status/Control Optional Concentration Alarm Status/ResetAlarm Status LED Color Alarm State Front Panel Menus Status LEDsMenu Navigation Basics Status LED Normal StateEditing Numeric Values Front Panel Menu Diagram OFF Point 5V Output MA Output Diag MenuAnalog Step-Test Signal Name Rear Panel Label Function Signal I/OZero Calibration Memory ResetSpan Calibration Span Cal Menu Flow Calibration Optional Electronic Flowmeter onlyFlow Cal Menu View Menu Config MenuParameter Meaning UnitsCount SETUP-VARS MenuProperty Meaning Default Range Value SETUP-STREAM MenuMulti-Stream Operation Digital Communications Overview Changing Com Port Hardware ConfigurationDigital Communications RS232 Configuration Property Value RS-232 ConfigurationFunction DB-9 General I/O Connector Pin SW2 DTE SW2 DCE Function DB-9 General I/O Connector Pin SW2 DCE RS-485 ConfigurationHardware Setup for Configuring the Ethernet Module Ethernet ConfigurationVerifying Network Hardware Connection Link LED Left Side Activity LED Right Side Color MeaningManually Assigning a Static IP Address Determining a Dynamically Assigned IP Address Dhcp ModeModbus Register Maps Modbus Commands SupportedDescription Function CodeDecimal Address Description Register Type Start Bit # Bits Real-Time Concentration and Status Registers Read OnlyInstrument Setup and Configuration Registers Read / Write Decimal Address Description Register Type Start Bit # Bits This page intentionally left blank Maintenance Item Recommended Interval Section Maintenance and AdjustmentsMaintenance Schedule Instrument Layout Instrument Layout, Rack Mount ConfigurationReplacing Internal Particulate Filter Instrument Layout, Nema ConfigurationUV Lamp Adjustment Sample Pump Replacement UV Lamp Replacement05509F DCN6409 Sensor Detail Sensor Valve ReplacementConfiguring the Analog Output Water Drop-Out Filter Option MaintenanceBoot Loader Operation Cleaning Exterior Surfaces of the M465LUpload Firmware Boot Loader Hardware SetupBoot Loader Software Setup Verify Firmware This page intentionally left blank Reference Drawings TroubleshootingPneumatic Diagram RACK-MOUNT Version only Interconnect DiagramStatus LED/Output Definitions Sensor OK Pneumatic ErrorInvalid Reading Check LampPressure Too Low Pressure Too HighThis page intentionally left blank Zero Calibration CalibrationFactory Calibration This page intentionally left blank Spare Parts 12.1. M465L Rack Mount Spare Parts ListPart Number Description 12.2. M465L Nema Spare Parts ListCONNECTOR, 16 PIN, W/SCREW Flange This page intentionally left blank Permeation Gas Dryer Index