9.5.UV Lamp Replacement

CAUTION

GENERAL SAFETY HAZARD

Do not look at the UV lamp while the unit is operating. UV light can cause eye damage. Always use safety glasses made from UV blocking material. (Generic plastic glasses are not adequate).

WARNING - RISK OF ELECTRICAL SHOCK

HIGH VOLTAGE MAY BE PRESENT WHEN POWER IS CONNECTED TO

THE INSTRUMENT!

1.Disconnect power from the M465L.

2.Remove the six screws and the top cover (Rack Mount Configuration) or open front panel (NEMA Configuration.)

3.Loosen the two UV lamp setscrews on the UV lamp housing (see Figure 9-1 or Figure 9-2 above.)

4.Unplug the lamp power cord from the connector labeled P1 on the sensor module.

5.Loosen the four slotted captive screws attaching the sensor module to the chassis.

6.Carefully slide the lamp out of housing. The sensor module may need to be raised slightly to completely remove the lamp.

7.Install the new lamp, seating it in the lamp housing until it stops.

8.Re-attach the sensor module captive screws to the chassis.

9.Re-tighten the two UV lamp setscrews.

10.Plug the lamp power cord into P1 on the sensor module.

11.Reconnect power to the instrument and turn on power switch. Let instrument warm up for at least 20 minutes.

12.Perform UV lamp adjustment procedure per Section 9.4.

9.6.Sample Pump Replacement

1.Disconnect power from the M465L.

2.Remove the six screws and the top cover (Rack Mount Configuration) or open front panel (NEMA Configuration.)

3.Locate sample pump assembly (See Figure 9-1.)

4.Cut off the clamps holding the tubing to the pump nipples and remove tubing. Note which tubing is connected to inlet and outlet.

70

05509F DCN6409

Image 72
Contents Model 465L Ozone Monitor Page About Teledyne Advanced Pollution Instrumentation Tapi This page intentionally left blank Important Safety Information General Safety HazardConsignes DE Sécurité Mise EN GardeAbout this Manual Part Name/Description NumberRevision History Table of Contents Digital Communications Figures Sensor Detail Tables Xii 05509F DCN6409 Introduction PrefaceThis page intentionally left blank Specifications and Agency Approvals SpecificationsSafety Multi-Stream SpecificationsApprovals EMCOther Type Certifications PhoneThis page intentionally left blank Principle of Operation = − InHg × ln ×  27305509F DCN6409 Getting Started Nema AC Power Connection to Monitor Pneumatic Connections, Rack Mount Configuration Pneumatic ConnectionsPneumatic Connections, Nema Configuration, Multi-Stream Sample Gas ConnectionSample Filtration Sample System Design in Humid Environments05509F DCN6409 Sample Conditioner Option Permeation Gas Dryer OptionExhaust Connection Rear Panel Electrical I/O ConnectionsLocation of I/O Connectors in the Nema Configuration Electrical I/O ConnectionsAnalog Output Status OutputsStatus Output Schematic Status Name Normal Operating Output # StateHi-Current Relay Outputs Relay # Function Normal Operating StateRelay # Function Normal Operating StateMulti-Stream Relay Outputs Optional Relay # Function MeaningDigital Serial Interfaces 6.1. RS232 Connection6.2. RS485 Connection Ethernet ConnectionHis page intentionally left blank Frequently Asked Questions This page intentionally left blank Options 05509F DCN6409 Operation Front Panel OverviewOzone Concentration Display Menu Navigation KeypadConcentration Alarm Status/Reset Alarm Status LED Color Alarm StateStream Selector Status/Control Optional Status LEDs Front Panel MenusMenu Navigation Basics Status LED Normal StateEditing Numeric Values Front Panel Menu Diagram OFF Diag Menu Analog Step-TestPoint 5V Output MA Output Signal I/O Signal Name Rear Panel Label FunctionMemory Reset Zero CalibrationSpan Calibration Flow Calibration Optional Electronic Flowmeter only Span Cal MenuFlow Cal Menu Config Menu View MenuParameter Meaning UnitsSETUP-VARS Menu CountSETUP-STREAM Menu Property Meaning Default Range ValueMulti-Stream Operation Changing Com Port Hardware Configuration Digital CommunicationsDigital Communications Overview RS232 Configuration RS-232 Configuration Function DB-9 General I/O Connector Pin SW2 DTE SW2 DCEProperty Value RS-485 Configuration Function DB-9 General I/O Connector Pin SW2 DCEEthernet Configuration Hardware Setup for Configuring the Ethernet ModuleVerifying Network Hardware Connection Link LED Left Side Activity LED Right Side Color MeaningDetermining a Dynamically Assigned IP Address Dhcp Mode Manually Assigning a Static IP AddressModbus Commands Supported Modbus Register MapsDescription Function CodeReal-Time Concentration and Status Registers Read Only Decimal Address Description Register Type Start Bit # BitsInstrument Setup and Configuration Registers Read / Write Decimal Address Description Register Type Start Bit # Bits This page intentionally left blank Maintenance and Adjustments Maintenance ScheduleMaintenance Item Recommended Interval Section Instrument Layout, Rack Mount Configuration Instrument LayoutInstrument Layout, Nema Configuration Replacing Internal Particulate FilterUV Lamp Adjustment UV Lamp Replacement Sample Pump Replacement05509F DCN6409 Sensor Valve Replacement Sensor DetailWater Drop-Out Filter Option Maintenance Configuring the Analog OutputCleaning Exterior Surfaces of the M465L Boot Loader OperationBoot Loader Hardware Setup Boot Loader Software SetupUpload Firmware Verify Firmware This page intentionally left blank Troubleshooting Pneumatic DiagramReference Drawings Interconnect Diagram RACK-MOUNT Version onlyStatus LED/Output Definitions Pneumatic Error Sensor OKInvalid Reading Check LampPressure Too High Pressure Too LowThis page intentionally left blank Calibration Factory CalibrationZero Calibration This page intentionally left blank 12.1. M465L Rack Mount Spare Parts List Spare Parts12.2. M465L Nema Spare Parts List Part Number DescriptionCONNECTOR, 16 PIN, W/SCREW Flange This page intentionally left blank Index Permeation Gas Dryer