Teledyne API Model 465L O3 Monitor Operation Manual
9.8.Water Drop-Out Filter (Option) Maintenance
Located externally on the right side of the enclosure, the Water
1.Disconnect the ozone supply line from the instrument.
2.Shut off flow to the instrument using the needle valve on the front panel.
3.Remove cap from the fitting located on the bottom of the filter and allow it to drain.
4.Replace cap and adjust flow to instrument.
Replacement of the particulate filter, a Teflon coalescing membrane should be rarely required. However, if it does become fouled, possibly indicated by a significant reduction in the gas pressure measured in the O3 cell, which would trigger a warning in the instrument, replace it as follows:
1.Remove the orange cap from the top of the water
2.Located the membrane on the underside of the cap assembly and replace it.
3.Tightly secure the orange cap back onto the water
9.9.Configuring the Analog Output
The analog output can be configured for either
1.Disconnect power from the M465L.
2.Remove the six screws and the top cover (Rack Mount Configuration) or open front panel (NEMA Configuration.)
3.Locate the Mainboard PCA (see Figure
4.Set the desired operation as shown in Figure
5.
05509F DCN6409 | 73 |