Teledyne 465L Water Drop-Out Filter Option Maintenance, Configuring the Analog Output

Page 75

Teledyne API Model 465L O3 Monitor Operation Manual

9.8.Water Drop-Out Filter (Option) Maintenance

Located externally on the right side of the enclosure, the Water Drop-Out Filter (Section 4.1.1.2) should be checked periodically for water build-up and drained. To drain this filter:

1.Disconnect the ozone supply line from the instrument.

2.Shut off flow to the instrument using the needle valve on the front panel.

3.Remove cap from the fitting located on the bottom of the filter and allow it to drain.

4.Replace cap and adjust flow to instrument.

Replacement of the particulate filter, a Teflon coalescing membrane should be rarely required. However, if it does become fouled, possibly indicated by a significant reduction in the gas pressure measured in the O3 cell, which would trigger a warning in the instrument, replace it as follows:

1.Remove the orange cap from the top of the water drop-out filter.

2.Located the membrane on the underside of the cap assembly and replace it.

3.Tightly secure the orange cap back onto the water drop-out filter.

9.9.Configuring the Analog Output

The analog output can be configured for either 0-5V DC or 4-20mA operation. To change or verify the configuration of the analog output:

1.Disconnect power from the M465L.

2.Remove the six screws and the top cover (Rack Mount Configuration) or open front panel (NEMA Configuration.)

3.Locate the Mainboard PCA (see Figure 9-1 or Figure 9-2.)

4.Set the desired operation as shown in Figure 9-4.

5.Re-Install the top cover or re-secure the front panel (NEMA Configuration.)

05509F DCN6409

73

Image 75
Contents Model 465L Ozone Monitor Page About Teledyne Advanced Pollution Instrumentation Tapi This page intentionally left blank General Safety Hazard Important Safety InformationMise EN Garde Consignes DE SécuritéPart Name/Description Number About this ManualRevision History Table of Contents Digital Communications Figures Sensor Detail Tables Xii 05509F DCN6409 Preface IntroductionThis page intentionally left blank Specifications Specifications and Agency ApprovalsEMC SafetyMulti-Stream Specifications ApprovalsPhone Other Type CertificationsThis page intentionally left blank = − InHg × ln ×  273 Principle of Operation05509F DCN6409 Getting Started Nema AC Power Connection to Monitor Pneumatic Connections Pneumatic Connections, Rack Mount ConfigurationSample Gas Connection Pneumatic Connections, Nema Configuration, Multi-StreamSample System Design in Humid Environments Sample Filtration05509F DCN6409 Permeation Gas Dryer Option Sample Conditioner OptionRear Panel Electrical I/O Connections Exhaust ConnectionStatus Outputs Location of I/O Connectors in the Nema ConfigurationElectrical I/O Connections Analog OutputStatus Name Normal Operating Output # State Status Output SchematicNormal Operating State Hi-Current Relay OutputsRelay # Function Normal Operating State Relay # FunctionRelay # Function Meaning Multi-Stream Relay Outputs OptionalEthernet Connection Digital Serial Interfaces6.1. RS232 Connection 6.2. RS485 ConnectionHis page intentionally left blank Frequently Asked Questions This page intentionally left blank Options 05509F DCN6409 Menu Navigation Keypad OperationFront Panel Overview Ozone Concentration DisplayConcentration Alarm Status/Reset Alarm Status LED Color Alarm StateStream Selector Status/Control Optional Status LED Normal State Status LEDsFront Panel Menus Menu Navigation BasicsEditing Numeric Values Front Panel Menu Diagram OFF Diag Menu Analog Step-TestPoint 5V Output MA Output Signal Name Rear Panel Label Function Signal I/OZero Calibration Memory ResetSpan Calibration Span Cal Menu Flow Calibration Optional Electronic Flowmeter onlyFlow Cal Menu Units Config MenuView Menu Parameter MeaningCount SETUP-VARS MenuProperty Meaning Default Range Value SETUP-STREAM MenuMulti-Stream Operation Changing Com Port Hardware Configuration Digital CommunicationsDigital Communications Overview RS232 Configuration RS-232 Configuration Function DB-9 General I/O Connector Pin SW2 DTE SW2 DCEProperty Value Function DB-9 General I/O Connector Pin SW2 DCE RS-485 ConfigurationLink LED Left Side Activity LED Right Side Color Meaning Ethernet ConfigurationHardware Setup for Configuring the Ethernet Module Verifying Network Hardware ConnectionManually Assigning a Static IP Address Determining a Dynamically Assigned IP Address Dhcp ModeFunction Code Modbus Commands SupportedModbus Register Maps DescriptionDecimal Address Description Register Type Start Bit # Bits Real-Time Concentration and Status Registers Read OnlyInstrument Setup and Configuration Registers Read / Write Decimal Address Description Register Type Start Bit # Bits This page intentionally left blank Maintenance and Adjustments Maintenance ScheduleMaintenance Item Recommended Interval Section Instrument Layout Instrument Layout, Rack Mount ConfigurationReplacing Internal Particulate Filter Instrument Layout, Nema ConfigurationUV Lamp Adjustment Sample Pump Replacement UV Lamp Replacement05509F DCN6409 Sensor Detail Sensor Valve ReplacementConfiguring the Analog Output Water Drop-Out Filter Option MaintenanceBoot Loader Operation Cleaning Exterior Surfaces of the M465LBoot Loader Hardware Setup Boot Loader Software SetupUpload Firmware Verify Firmware This page intentionally left blank Troubleshooting Pneumatic DiagramReference Drawings RACK-MOUNT Version only Interconnect DiagramStatus LED/Output Definitions Check Lamp Pneumatic ErrorSensor OK Invalid ReadingPressure Too Low Pressure Too HighThis page intentionally left blank Calibration Factory CalibrationZero Calibration This page intentionally left blank Spare Parts 12.1. M465L Rack Mount Spare Parts ListPart Number Description 12.2. M465L Nema Spare Parts ListCONNECTOR, 16 PIN, W/SCREW Flange This page intentionally left blank Permeation Gas Dryer Index