7.2.6.SETUP-VARS Menu

The VARS menu allows viewing and editing of various global setup variables that effect how the M465L operates. These variables are stored in the instrument’s non-volatile memory. Table 7-7 lists these variables and their function.

Table 7-7: VARS List

 

VAR

 

 

Meaning

 

Default

 

Range

 

 

 

 

 

 

 

Value

 

 

 

 

S/N

 

 

Serial number of M465L. This number should match

N/A

 

1 – 9999

 

 

 

 

 

 

 

 

 

the serial number tag on the rear of the instrument.

 

 

 

 

 

ALARM

 

 

Controls the latching behavior of the concentration alarms.

 

LATCHING,

 

MODE

 

 

See Section 7.1.4 for additional details.

 

NON-LATCHING

 

STREAM

 

 

 

 

1

1,3,6

 

 

 

 

Specifies the number of streams available in the instrument. Note that the

 

 

COUNT*

 

 

hardware in the instrument must be also configured to support the number

 

 

 

 

 

 

 

 

of streams specified.

 

 

 

 

 

DWEL

 

 

Dwell value for stream selector operation. See Section 7.3

30.0

 

0.0 – 30.0

 

 

 

 

for additional details.

Seconds

 

Seconds

 

 

 

 

Multi-stream units with the Sample Conditioner option

45.0

 

0.0 – 45.0

 

 

 

 

(Section 4.1.1.2), require additional dwell time to purge the

Seconds

 

Seconds

 

 

 

 

filter between stream measurements.

 

 

 

 

 

ALARM

 

 

Number of readings above alarm limit required to trigger

5

1 - 10

 

 

DELAY

 

 

alarms

 

 

 

 

MAN MODE

 

 

Enables/disables the Manual mode timeout feature. See

ON

 

ON, OFF

TIMEOUT

 

 

Section 7.1.3 for more detail.

 

 

 

 

 

BC FILT

 

 

Boxcar filter length for concentration filter. Increasing the

10

1-30

 

 

SIZE*

 

 

size of the filter will smooth out the concentration readings

 

 

 

 

 

 

 

 

but will also increase the response time.

 

 

 

 

 

FLOW

 

 

Enables/disables operation of optional electronic flowmeter

OFF

ON, OFF

 

ENABLE

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

COMM

 

 

MODBUS address for serial communications

1

1 - 20

 

 

ADDRESS

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

UNIT

 

 

Concentration unit-of-measure setting.

PPM

PPM, PPB, UGM

 

 

 

 

 

 

 

 

 

SCROLL

 

 

Inter-character delay for the front panel display, to increase

20

6-100

 

 

DELAY

 

 

speed (lower value) or decrease speed (higher value) the

 

 

 

 

 

 

 

 

display scroll.

 

 

 

 

USER SLOPE

 

User-editable calibration slope value

1.000

 

0.8 – 1.0

 

 

 

 

 

 

 

 

 

UNITS

 

 

Standard (20ºC) or Normal (0ºC) temperature selection for

NORMAL

 

NORMAL

 

TEMP

 

 

g/Nm3 units of measure.

 

 

STANDARD

* Changing the value of these VARS forces the instrument to automatically reboot for the changes to take effect.

52

05509F DCN6409

Image 54
Contents Model 465L Ozone Monitor Page About Teledyne Advanced Pollution Instrumentation Tapi This page intentionally left blank Important Safety Information General Safety HazardConsignes DE Sécurité Mise EN GardeAbout this Manual Part Name/Description NumberRevision History Table of Contents Digital Communications Figures Sensor Detail Tables Xii 05509F DCN6409 Introduction PrefaceThis page intentionally left blank Specifications and Agency Approvals SpecificationsApprovals SafetyMulti-Stream Specifications EMCOther Type Certifications PhoneThis page intentionally left blank Principle of Operation = − InHg × ln ×  27305509F DCN6409 Getting Started Nema AC Power Connection to Monitor Pneumatic Connections, Rack Mount Configuration Pneumatic ConnectionsPneumatic Connections, Nema Configuration, Multi-Stream Sample Gas ConnectionSample Filtration Sample System Design in Humid Environments05509F DCN6409 Sample Conditioner Option Permeation Gas Dryer OptionExhaust Connection Rear Panel Electrical I/O ConnectionsAnalog Output Location of I/O Connectors in the Nema ConfigurationElectrical I/O Connections Status OutputsStatus Output Schematic Status Name Normal Operating Output # StateRelay # Function Hi-Current Relay OutputsRelay # Function Normal Operating State Normal Operating StateMulti-Stream Relay Outputs Optional Relay # Function Meaning6.2. RS485 Connection Digital Serial Interfaces6.1. RS232 Connection Ethernet ConnectionHis page intentionally left blank Frequently Asked Questions This page intentionally left blank Options 05509F DCN6409 Ozone Concentration Display OperationFront Panel Overview Menu Navigation KeypadConcentration Alarm Status/Reset Alarm Status LED Color Alarm StateStream Selector Status/Control Optional Menu Navigation Basics Status LEDsFront Panel Menus Status LED Normal StateEditing Numeric Values Front Panel Menu Diagram OFF Diag Menu Analog Step-TestPoint 5V Output MA Output Signal I/O Signal Name Rear Panel Label FunctionMemory Reset Zero CalibrationSpan Calibration Flow Calibration Optional Electronic Flowmeter only Span Cal MenuFlow Cal Menu Parameter Meaning Config MenuView Menu UnitsSETUP-VARS Menu CountSETUP-STREAM Menu Property Meaning Default Range ValueMulti-Stream Operation Changing Com Port Hardware Configuration Digital CommunicationsDigital Communications Overview RS232 Configuration RS-232 Configuration Function DB-9 General I/O Connector Pin SW2 DTE SW2 DCEProperty Value RS-485 Configuration Function DB-9 General I/O Connector Pin SW2 DCEVerifying Network Hardware Connection Ethernet ConfigurationHardware Setup for Configuring the Ethernet Module Link LED Left Side Activity LED Right Side Color MeaningDetermining a Dynamically Assigned IP Address Dhcp Mode Manually Assigning a Static IP AddressDescription Modbus Commands SupportedModbus Register Maps Function CodeReal-Time Concentration and Status Registers Read Only Decimal Address Description Register Type Start Bit # BitsInstrument Setup and Configuration Registers Read / Write Decimal Address Description Register Type Start Bit # Bits This page intentionally left blank Maintenance and Adjustments Maintenance ScheduleMaintenance Item Recommended Interval Section Instrument Layout, Rack Mount Configuration Instrument LayoutInstrument Layout, Nema Configuration Replacing Internal Particulate FilterUV Lamp Adjustment UV Lamp Replacement Sample Pump Replacement05509F DCN6409 Sensor Valve Replacement Sensor DetailWater Drop-Out Filter Option Maintenance Configuring the Analog OutputCleaning Exterior Surfaces of the M465L Boot Loader OperationBoot Loader Hardware Setup Boot Loader Software SetupUpload Firmware Verify Firmware This page intentionally left blank Troubleshooting Pneumatic DiagramReference Drawings Interconnect Diagram RACK-MOUNT Version onlyStatus LED/Output Definitions Invalid Reading Pneumatic ErrorSensor OK Check LampPressure Too High Pressure Too LowThis page intentionally left blank Calibration Factory CalibrationZero Calibration This page intentionally left blank 12.1. M465L Rack Mount Spare Parts List Spare Parts12.2. M465L Nema Spare Parts List Part Number DescriptionCONNECTOR, 16 PIN, W/SCREW Flange This page intentionally left blank Index Permeation Gas Dryer