Teledyne 465L operation manual Flow Calibration Optional Electronic Flowmeter only, Span Cal Menu

Page 51

Teledyne API Model 465L O3 Monitor Operation Manual

Enter Span target

concentration (actual ozone concentration being supplied to monitor)

SPAN CAL ENT

900

ENT

ENT TO

ENT

SUCCESSFUL?

Yes

RETURN TO

SPAN

CONC MODE

 

 

 

 

 

 

 

 

 

CFG

 

 

 

 

 

 

 

 

 

No

 

 

 

 

 

 

 

OUT OF

 

 

 

 

 

 

 

RANGE

 

 

 

 

 

 

 

ENT

 

 

Figure 7-4: Span Cal Menu

7.2.3.6. Flow Calibration (Optional Electronic Flowmeter only)

The flow calibration allows the instrument to calibrate its internal flowmeter reading. This should be done with a calibrated flowmeter connected to the “Sample In” port of the M465L. Figure 7-5 below shows the Flow Cal menu. After the instrument has stabilized on the source of span gas, navigate to the Flow Cal menu (See

Figure 7-2) and Press ENT. Next enter the Flow Cal target (the actual flow as measured at the sample inlet) and press ENT. The display will next show a confirmation menu, ENT TO CAL. Press ENT to perform the Flow Calibration, or CFG to abort back to the start. If the calibration is successful, the display will return to the concentration. If the calibration cannot be performed, an OUT OF RANGE error will be displayed. Press ENT to confirm and the display will return to the start of the Flow Cal menu.

If the OUT OF RANGE error occurs, it means the Flow Cal cannot be performed because it would result in an out of range flow slope value for the monitor. This means that either the flow sensor in the monitor is malfunctioning, causing improper readings, or the actual flow is different than the target value being entered.

05509F DCN6409

49

Image 51
Contents Model 465L Ozone Monitor Page About Teledyne Advanced Pollution Instrumentation Tapi This page intentionally left blank General Safety Hazard Important Safety InformationMise EN Garde Consignes DE SécuritéPart Name/Description Number About this ManualRevision History Table of Contents Digital Communications Figures Sensor Detail Tables Xii 05509F DCN6409 Preface IntroductionThis page intentionally left blank Specifications Specifications and Agency ApprovalsEMC SafetyMulti-Stream Specifications ApprovalsPhone Other Type CertificationsThis page intentionally left blank = − InHg × ln ×  273 Principle of Operation05509F DCN6409 Getting Started Nema AC Power Connection to Monitor Pneumatic Connections Pneumatic Connections, Rack Mount ConfigurationSample Gas Connection Pneumatic Connections, Nema Configuration, Multi-StreamSample System Design in Humid Environments Sample Filtration05509F DCN6409 Permeation Gas Dryer Option Sample Conditioner OptionRear Panel Electrical I/O Connections Exhaust ConnectionStatus Outputs Location of I/O Connectors in the Nema ConfigurationElectrical I/O Connections Analog OutputStatus Name Normal Operating Output # State Status Output SchematicNormal Operating State Hi-Current Relay OutputsRelay # Function Normal Operating State Relay # FunctionRelay # Function Meaning Multi-Stream Relay Outputs OptionalEthernet Connection Digital Serial Interfaces6.1. RS232 Connection 6.2. RS485 ConnectionHis page intentionally left blank Frequently Asked Questions This page intentionally left blank Options 05509F DCN6409 Menu Navigation Keypad OperationFront Panel Overview Ozone Concentration DisplayConcentration Alarm Status/Reset Alarm Status LED Color Alarm StateStream Selector Status/Control Optional Status LED Normal State Status LEDsFront Panel Menus Menu Navigation BasicsEditing Numeric Values Front Panel Menu Diagram OFF Diag Menu Analog Step-TestPoint 5V Output MA Output Signal Name Rear Panel Label Function Signal I/OZero Calibration Memory ResetSpan Calibration Span Cal Menu Flow Calibration Optional Electronic Flowmeter onlyFlow Cal Menu Units Config MenuView Menu Parameter MeaningCount SETUP-VARS MenuProperty Meaning Default Range Value SETUP-STREAM MenuMulti-Stream Operation Changing Com Port Hardware Configuration Digital CommunicationsDigital Communications Overview RS232 Configuration RS-232 Configuration Function DB-9 General I/O Connector Pin SW2 DTE SW2 DCEProperty Value Function DB-9 General I/O Connector Pin SW2 DCE RS-485 ConfigurationLink LED Left Side Activity LED Right Side Color Meaning Ethernet ConfigurationHardware Setup for Configuring the Ethernet Module Verifying Network Hardware ConnectionManually Assigning a Static IP Address Determining a Dynamically Assigned IP Address Dhcp ModeFunction Code Modbus Commands SupportedModbus Register Maps DescriptionDecimal Address Description Register Type Start Bit # Bits Real-Time Concentration and Status Registers Read OnlyInstrument Setup and Configuration Registers Read / Write Decimal Address Description Register Type Start Bit # Bits This page intentionally left blank Maintenance and Adjustments Maintenance ScheduleMaintenance Item Recommended Interval Section Instrument Layout Instrument Layout, Rack Mount ConfigurationReplacing Internal Particulate Filter Instrument Layout, Nema ConfigurationUV Lamp Adjustment Sample Pump Replacement UV Lamp Replacement05509F DCN6409 Sensor Detail Sensor Valve ReplacementConfiguring the Analog Output Water Drop-Out Filter Option MaintenanceBoot Loader Operation Cleaning Exterior Surfaces of the M465LBoot Loader Hardware Setup Boot Loader Software SetupUpload Firmware Verify Firmware This page intentionally left blank Troubleshooting Pneumatic DiagramReference Drawings RACK-MOUNT Version only Interconnect DiagramStatus LED/Output Definitions Check Lamp Pneumatic ErrorSensor OK Invalid ReadingPressure Too Low Pressure Too HighThis page intentionally left blank Calibration Factory CalibrationZero Calibration This page intentionally left blank Spare Parts 12.1. M465L Rack Mount Spare Parts ListPart Number Description 12.2. M465L Nema Spare Parts ListCONNECTOR, 16 PIN, W/SCREW Flange This page intentionally left blank Permeation Gas Dryer Index