Teledyne API Model 465L O3 Monitor Operation Manual
The firmware loader software should display a message “Boot loader found, uploading…”. If this message is not displayed, power down the instrument and receck the COM port configuration and cable (see Section 9.11.1) and retry.
3.When the upload is complete, a message,
NOTE
It is not recommended to stop the writing process half way through. This will cause the firmware to not be fully uploaded.
9.11.2.2. Verify Firmware
The verify function compares the instrument’s current firmware to the selected HEX file (see Section 9.11.2, step 5) and confirms whether or not the files are a match.
1.Press “Verify”. The screen will display, “Wait for boot loader”.
2.When the screen displays, “Power cycle instrument now”, power up the instrument.
NOTE
Search for boot loader will time out in 20 seconds. If the instrument is not powered up in 20 seconds, the screen will display, “No boot loader detected.” Go back to step 1 to verify the firmware.
The firmware loader software should display a message “Boot loader found, verifying…”. If this message is not displayed, power down the instrument and receck the COM port configuration and cable (see Section 9.11.1) and retry.
3.When the upload is complete, a message,
NOTE
A verification process can be stopped at any point in time. This will not affect the firmware that is already uploaded on the instrument. The instrument will automatically go to the firmware in 10 seconds.
76 | 05509F DCN6409 |