Teledyne API Model 465L O3 Monitor Operation Manual

The firmware loader software should display a message “Boot loader found, uploading…”. If this message is not displayed, power down the instrument and receck the COM port configuration and cable (see Section 9.11.1) and retry.

3.When the upload is complete, a message, “Write-flash completed successfully” should be displayed. If it is an unsuccessful upload, a message, “Write-flash error” is displayed. If a Write-flash error occurs, ensure that the correct PC COM port is selected for the instrument (see Section 9.11.2, step 4).

NOTE

It is not recommended to stop the writing process half way through. This will cause the firmware to not be fully uploaded.

9.11.2.2. Verify Firmware

The verify function compares the instrument’s current firmware to the selected HEX file (see Section 9.11.2, step 5) and confirms whether or not the files are a match.

1.Press “Verify”. The screen will display, “Wait for boot loader”.

2.When the screen displays, “Power cycle instrument now”, power up the instrument.

NOTE

Search for boot loader will time out in 20 seconds. If the instrument is not powered up in 20 seconds, the screen will display, “No boot loader detected.” Go back to step 1 to verify the firmware.

The firmware loader software should display a message “Boot loader found, verifying…”. If this message is not displayed, power down the instrument and receck the COM port configuration and cable (see Section 9.11.1) and retry.

3.When the upload is complete, a message, “Verify-flash completed successfully” should be displayed. If it is an unsuccessful upload, a message “Verify-flash error” is displayed. If a Verify-flash error occurs, ensure that the correct PC COM port is selected for the instrument (see Section 9.11.2, step 4).

NOTE

A verification process can be stopped at any point in time. This will not affect the firmware that is already uploaded on the instrument. The instrument will automatically go to the firmware in 10 seconds.

76

05509F DCN6409

Image 78
Contents Model 465L Ozone Monitor Page About Teledyne Advanced Pollution Instrumentation Tapi This page intentionally left blank Important Safety Information General Safety HazardConsignes DE Sécurité Mise EN GardeAbout this Manual Part Name/Description NumberRevision History Table of Contents Digital Communications Figures Sensor Detail Tables Xii 05509F DCN6409 Introduction PrefaceThis page intentionally left blank Specifications and Agency Approvals SpecificationsApprovals SafetyMulti-Stream Specifications EMCOther Type Certifications PhoneThis page intentionally left blank Principle of Operation = − InHg × ln ×  27305509F DCN6409 Getting Started Nema AC Power Connection to Monitor Pneumatic Connections, Rack Mount Configuration Pneumatic ConnectionsPneumatic Connections, Nema Configuration, Multi-Stream Sample Gas ConnectionSample Filtration Sample System Design in Humid Environments05509F DCN6409 Sample Conditioner Option Permeation Gas Dryer OptionExhaust Connection Rear Panel Electrical I/O ConnectionsAnalog Output Location of I/O Connectors in the Nema ConfigurationElectrical I/O Connections Status OutputsStatus Output Schematic Status Name Normal Operating Output # StateRelay # Function Hi-Current Relay OutputsRelay # Function Normal Operating State Normal Operating StateMulti-Stream Relay Outputs Optional Relay # Function Meaning6.2. RS485 Connection Digital Serial Interfaces6.1. RS232 Connection Ethernet ConnectionHis page intentionally left blank Frequently Asked Questions This page intentionally left blank Options 05509F DCN6409 Ozone Concentration Display OperationFront Panel Overview Menu Navigation KeypadConcentration Alarm Status/Reset Alarm Status LED Color Alarm StateStream Selector Status/Control Optional Menu Navigation Basics Status LEDsFront Panel Menus Status LED Normal StateEditing Numeric Values Front Panel Menu Diagram OFF Diag Menu Analog Step-TestPoint 5V Output MA Output Signal I/O Signal Name Rear Panel Label FunctionMemory Reset Zero CalibrationSpan Calibration Flow Calibration Optional Electronic Flowmeter only Span Cal MenuFlow Cal Menu Parameter Meaning Config MenuView Menu UnitsSETUP-VARS Menu CountSETUP-STREAM Menu Property Meaning Default Range ValueMulti-Stream Operation Changing Com Port Hardware Configuration Digital CommunicationsDigital Communications Overview RS232 Configuration RS-232 Configuration Function DB-9 General I/O Connector Pin SW2 DTE SW2 DCEProperty Value RS-485 Configuration Function DB-9 General I/O Connector Pin SW2 DCEVerifying Network Hardware Connection Ethernet ConfigurationHardware Setup for Configuring the Ethernet Module Link LED Left Side Activity LED Right Side Color MeaningDetermining a Dynamically Assigned IP Address Dhcp Mode Manually Assigning a Static IP AddressDescription Modbus Commands SupportedModbus Register Maps Function CodeReal-Time Concentration and Status Registers Read Only Decimal Address Description Register Type Start Bit # BitsInstrument Setup and Configuration Registers Read / Write Decimal Address Description Register Type Start Bit # Bits This page intentionally left blank Maintenance and Adjustments Maintenance ScheduleMaintenance Item Recommended Interval Section Instrument Layout, Rack Mount Configuration Instrument LayoutInstrument Layout, Nema Configuration Replacing Internal Particulate FilterUV Lamp Adjustment UV Lamp Replacement Sample Pump Replacement05509F DCN6409 Sensor Valve Replacement Sensor DetailWater Drop-Out Filter Option Maintenance Configuring the Analog OutputCleaning Exterior Surfaces of the M465L Boot Loader OperationBoot Loader Hardware Setup Boot Loader Software SetupUpload Firmware Verify Firmware This page intentionally left blank Troubleshooting Pneumatic DiagramReference Drawings Interconnect Diagram RACK-MOUNT Version onlyStatus LED/Output Definitions Invalid Reading Pneumatic ErrorSensor OK Check LampPressure Too High Pressure Too LowThis page intentionally left blank Calibration Factory CalibrationZero Calibration This page intentionally left blank 12.1. M465L Rack Mount Spare Parts List Spare Parts12.2. M465L Nema Spare Parts List Part Number DescriptionCONNECTOR, 16 PIN, W/SCREW Flange This page intentionally left blank Index Permeation Gas Dryer