Teledyne API Model 465L O3 Monitor Operation Manual

7.2.3.5. Span Calibration

The span calibration allows the instrument to calibrate its internal slope factor based on a known ozone concentration. This should be done with a source of ozone span gas connected to the “Sample In” port of the M465L. Allow the instrument to stabilize on the span gas source for a minimum of 1 hour before performing the span calibration.

NOTE

A Span Calibration should only be performed with a precision source of ozone calibration gas, such as a Teledyne API Model 700 or Model 703 calibrator. Simple ozone generators without a measurement feedback system should never be used for performing span calibrations.

If you are unsure regarding the suitability of a particular source of calibration gas, contact Customer Service at Teledyne API for assistance.

Figure 7-4 below shows the Span Cal menu. After the instrument has stabilized on the source of

span gas, navigate to the Span Cal menu (See

Figure 7-2) and Press ENT. Next enter the Span Target concentration (the actual concentration of ozone being supplied to the monitor) and press ENT. The display will next show a confirmation menu, ENT TO SPAN. Press ENT to perform the Span Calibration, or CFG to abort back to the start. If the calibration is successful, the display will return to the concentration menu and the monitor reading should change adjust to read very close to the target value. If the calibration cannot be performed, an OUT OF RANGE error will be displayed. Press ENT to confirm and the display will return to the start of the Span Cal menu.

If the OUT OF RANGE error occurs, it means the Span Cal cannot be performed because it would result in an out of range slope value for the monitor. This means that either the sensor in the monitor is malfunctioning, causing improper readings, or the actual ozone concentration being supplied to the monitor is different than the target value being entered.

48

05509F DCN6409

Image 50
Contents Model 465L Ozone Monitor Page About Teledyne Advanced Pollution Instrumentation Tapi This page intentionally left blank Important Safety Information General Safety HazardConsignes DE Sécurité Mise EN GardeAbout this Manual Part Name/Description NumberRevision History Table of Contents Digital Communications Figures Sensor Detail Tables Xii 05509F DCN6409 Introduction PrefaceThis page intentionally left blank Specifications and Agency Approvals SpecificationsApprovals SafetyMulti-Stream Specifications EMCOther Type Certifications PhoneThis page intentionally left blank Principle of Operation = − InHg × ln ×  27305509F DCN6409 Getting Started Nema AC Power Connection to Monitor Pneumatic Connections, Rack Mount Configuration Pneumatic ConnectionsPneumatic Connections, Nema Configuration, Multi-Stream Sample Gas ConnectionSample Filtration Sample System Design in Humid Environments05509F DCN6409 Sample Conditioner Option Permeation Gas Dryer OptionExhaust Connection Rear Panel Electrical I/O ConnectionsAnalog Output Location of I/O Connectors in the Nema ConfigurationElectrical I/O Connections Status OutputsStatus Output Schematic Status Name Normal Operating Output # StateRelay # Function Hi-Current Relay OutputsRelay # Function Normal Operating State Normal Operating StateMulti-Stream Relay Outputs Optional Relay # Function Meaning6.2. RS485 Connection Digital Serial Interfaces6.1. RS232 Connection Ethernet ConnectionHis page intentionally left blank Frequently Asked Questions This page intentionally left blank Options 05509F DCN6409 Ozone Concentration Display OperationFront Panel Overview Menu Navigation KeypadStream Selector Status/Control Optional Concentration Alarm Status/ResetAlarm Status LED Color Alarm State Menu Navigation Basics Status LEDsFront Panel Menus Status LED Normal StateEditing Numeric Values Front Panel Menu Diagram OFF Point 5V Output MA Output Diag MenuAnalog Step-Test Signal I/O Signal Name Rear Panel Label FunctionMemory Reset Zero CalibrationSpan Calibration Flow Calibration Optional Electronic Flowmeter only Span Cal MenuFlow Cal Menu Parameter Meaning Config MenuView Menu UnitsSETUP-VARS Menu CountSETUP-STREAM Menu Property Meaning Default Range ValueMulti-Stream Operation Digital Communications Overview Changing Com Port Hardware ConfigurationDigital Communications RS232 Configuration Property Value RS-232 ConfigurationFunction DB-9 General I/O Connector Pin SW2 DTE SW2 DCE RS-485 Configuration Function DB-9 General I/O Connector Pin SW2 DCEVerifying Network Hardware Connection Ethernet ConfigurationHardware Setup for Configuring the Ethernet Module Link LED Left Side Activity LED Right Side Color MeaningDetermining a Dynamically Assigned IP Address Dhcp Mode Manually Assigning a Static IP AddressDescription Modbus Commands SupportedModbus Register Maps Function CodeReal-Time Concentration and Status Registers Read Only Decimal Address Description Register Type Start Bit # BitsInstrument Setup and Configuration Registers Read / Write Decimal Address Description Register Type Start Bit # Bits This page intentionally left blank Maintenance Item Recommended Interval Section Maintenance and AdjustmentsMaintenance Schedule Instrument Layout, Rack Mount Configuration Instrument LayoutInstrument Layout, Nema Configuration Replacing Internal Particulate FilterUV Lamp Adjustment UV Lamp Replacement Sample Pump Replacement05509F DCN6409 Sensor Valve Replacement Sensor DetailWater Drop-Out Filter Option Maintenance Configuring the Analog OutputCleaning Exterior Surfaces of the M465L Boot Loader OperationUpload Firmware Boot Loader Hardware SetupBoot Loader Software Setup Verify Firmware This page intentionally left blank Reference Drawings TroubleshootingPneumatic Diagram Interconnect Diagram RACK-MOUNT Version onlyStatus LED/Output Definitions Invalid Reading Pneumatic ErrorSensor OK Check LampPressure Too High Pressure Too LowThis page intentionally left blank Zero Calibration CalibrationFactory Calibration This page intentionally left blank 12.1. M465L Rack Mount Spare Parts List Spare Parts12.2. M465L Nema Spare Parts List Part Number DescriptionCONNECTOR, 16 PIN, W/SCREW Flange This page intentionally left blank Index Permeation Gas Dryer