Teledyne API Model 465L O3 Monitor Operation Manual
7.2.7.SETUP-STREAM Menu
The STREAM menu allows the operator to view and edit properties of each measurement stream available. For each stream the following properties are available:
Table
Property | Meaning | Default | Range |
|
| Value |
|
DURATION | Stream measurement duration for | 60.0 Seconds | 60.0 – 360.0 |
| AUTO |
| Seconds |
| See Section 7.3 for additional |
|
|
| details. |
|
|
HI ALARM | HI alarm trigger value | 0.100 PPM | .010 – 9.999 |
|
|
| PPM* |
HIHI ALARM | 0.300 PPM | .010 – 9.999 | |
|
|
| PPM* |
ON/OFF HI | HI alarm enable, enables HI alarm | ON | ON/OFF |
ALARM | operation |
|
|
ON/OFF HI HI | ON | ON/OFF | |
ALARM | alarm operation |
|
|
RANG | Concentration range for specified | 1.0 PPM | 1.0 – 500.0 |
| stream. Used to scale the analog |
| PPM |
| output. |
|
|
ENABLE | Allows the choice of viewing or | ON |
|
| skipping each stream in a multi- |
|
|
| channel instrument when cycling |
|
|
| through the streams in Auto or |
|
|
| Manual modes. OFF = skip. At |
|
|
| least one stream must be set to ON |
|
|
| (enabled). |
|
|
* For low concentration instrument ranges. For certain range configurations, the alarms may be set to higher values
05509F DCN6409 | 53 |