Teledyne SERIES 4030, HEATED THC manual Electrometer-Amplifier, Anode Power Supply

Page 15

Introduction

Figure 2-1: Flame Ionization Cell

The electrode section of the cell may be removed for inspection by turning off the power, disconnecting the electrode lead plug, and removing the screws, which retain the electrode assembly in the sensor body.

2.3.1 Electrometer-Amplifier

The collector cable is coupled directly to a coaxial fitting located on the electrometer-amplifier PC board. The PC board is located on the side panel next to but outside of the isothermal chamber. See Figure 2-1 and consists of an electrometer amplifier and an operational amplifier. This circuit is a very high-gain, current-to-voltage converter circuit, having an input impedance measuring in the billions of ohms. It is static sensitive and highly susceptible to contamination. Special care must be taken in handling this PC board.

2.3.2 Anode Power Supply

The high voltage anode power supply components are mounted on the anode power supply printed circuit board. High voltage regulation is achieved through the use of series-connected zener diodes. The simplicity of this circuit’s design can be attributed to the extremely low current demand of the anode circuit. The positive output voltage is

Teledyne Analytical Instruments

13

Image 15
Contents Series 4030 Heated THC Copyright 2004 Teledyne Analytical Instruments Safety Messages Series Additional Safety Information Series Procedure for Removal of Internal Inaccessible Shock Hazards Series Introduction Main Features of the AnalyzerAnalyzer Description Principle of OperationIntroduction ApplicationsInput Porting Sample SystemFlame Ionization Detection Cell Detection CellGas Flow Control System Fuel and Blanket Air SystemsIntroduction Electrometer-Amplifier Anode Power SupplyFlame Ignition Circuit Flame Guard CircuitMounting the Analyzer Unpacking the AnalyzerElectrical Power Connections User ConnectionsGas Connections Installation Electrical Connections Equipment Interface Connector Pin Arrangement Pin Function Threshold Alarm Pin Contact Pin Out Table Pin out of 50 pins D-Sub Connector4 RS-232 Port Commands via RS-232 InputSupporting Gases Required RS-232 OptionsSample Bypass Vent Carrier Gas Placing the System in OperationActivating the Support Gases 1 AirFuel Flame IgnitionSample Pump Analyzer OperationVerification of the Flame Guard Circuit Ignition and/or Flame Guard Circuit FailureOperation Style Conventions Default ParametersEnter Navigation and Data EntryEscape Line Menu StructureMain Menus Autocal is not available for standard configurations Logout Span ANALOG-OUT Adjust Adjusting the Gain Setting up an AUTO-CALPassword Protection Entering a Password Repeat Password Entry Screen Logging Out Model Screen System Self-Diagnostic TestSpan Alarms Function AL-2 Source SMP Range Function Changing Stream Change Stream is not available for standard configurationsAnalog output Adjustment Advanced User Functions StandbyTiming Qty Description Recommended Spare Parts ListAnalytical Instruments Drawing List