Operation

Operation

4.1Placing the System in Operation

1.Plug the power in the AC power inlet in the back of the Analyzer and the Sample Selector Unit.

2.Allow at least 2 hours warm-up (heat up sensor & sample system) after making the air adjustment described below. Warm up time is set by the software at the factory but may be overridden by pressing the <ENTER> key.

DO NOT attempt to ignite the flame during warm up countdown. Condensation will occur.

4.2Activating the Support Gases

4.2.1Air

1.Set the air tank regulator to 30 psig.

2.Adjust the Analyzer Air regulator until the air pressure gauge reads the recommended air pressure of 9 psig.

After the air is flowing through the sensor and warm-up time has been completed, activate the following gases:

4.2.2 Carrier Gas

Set the Carrier gas tank regulator to 50 psig and adjust the Analyzer Sample regulator until its (sample) pressure gauge reads the recommended sample pressure of 9.5 psig.

4.2.3 Span Gas

The Span gas containing certified mixture of methane in air, recommended range 6 to 8 ppm methane in zero grade air, may be connected to the Span Manifold or the ¼ inch connection on the back of the Sample Selector Module.

Teledyne Analytical Instruments

27

Image 29
Contents Series 4030 Heated THC Copyright 2004 Teledyne Analytical Instruments Safety Messages Series Additional Safety Information Series Procedure for Removal of Internal Inaccessible Shock Hazards Series Introduction Main Features of the AnalyzerAnalyzer Description Principle of OperationIntroduction ApplicationsInput Porting Sample SystemGas Flow Control System Detection CellFuel and Blanket Air Systems Flame Ionization Detection CellIntroduction Electrometer-Amplifier Anode Power SupplyFlame Ignition Circuit Flame Guard CircuitMounting the Analyzer Unpacking the AnalyzerGas Connections Electrical Power ConnectionsUser Connections Installation Electrical Connections Equipment Interface Connector Pin Arrangement Pin Function Threshold Alarm Pin Contact Pin Out Table Pin out of 50 pins D-Sub Connector4 RS-232 Port Commands via RS-232 InputSupporting Gases Required RS-232 OptionsSample Bypass Vent Activating the Support Gases Placing the System in Operation1 Air Carrier GasFuel Flame IgnitionVerification of the Flame Guard Circuit Analyzer OperationIgnition and/or Flame Guard Circuit Failure Sample PumpOperation Style Conventions Default ParametersEnter Navigation and Data EntryEscape Line Menu StructureMain Menus Autocal is not available for standard configurations Logout Span ANALOG-OUT Adjust Adjusting the Gain Setting up an AUTO-CALPassword Protection Entering a Password Repeat Password Entry Screen Logging Out Model Screen System Self-Diagnostic TestSpan Alarms Function AL-2 Source SMP Range Function Changing Stream Change Stream is not available for standard configurationsAnalog output Adjustment Timing Advanced User FunctionsStandby Qty Description Recommended Spare Parts ListAnalytical Instruments Drawing List