Teledyne HEATED THC, SERIES 4030 manual Pin Contact

Page 24

Installation

Table 3-2: Alarm Relay Contact Pins

Pin Contact

45Threshold Alarm 1, normally closed contact

28Threshold Alarm 1, moving contact

46Threshold Alarm 1, normally open contact

42Threshold Alarm 2, normally closed contact

44Threshold Alarm 2, moving contact

43Threshold Alarm 2, normally open contact

36System Alarm, normally closed contact

20System Alarm, moving contact

37System Alarm, normally open contact

3.3.3.6RANGE ID RELAYS

There are three dedicated Range ID relay contacts. They are assigned to relays in ascending order—Low range is assigned to Range 1 ID, Medium range is assigned to Range 2 ID, and High range is assigned to Range 3 ID. Table 3-6 lists the pin connections. Contacts are normally open, and they close when 4040 switches to that particular range.

Table 3-3: Range ID Relay Connections

Pin Function

21Range 1 ID Contact

38Range 1 ID Contact

22Range 2 ID Contact

39Range 2 ID Contact

19Range 3 ID Contact

18Range 3 ID Contact

34Not Used

35Not Used

Teledyne Analytical Instruments

22

Image 24
Contents Series 4030 Heated THC Copyright 2004 Teledyne Analytical Instruments Safety Messages Series Additional Safety Information Series Procedure for Removal of Internal Inaccessible Shock Hazards Series Main Features of the Analyzer IntroductionPrinciple of Operation Analyzer DescriptionApplications IntroductionSample System Input PortingDetection Cell Gas Flow Control SystemFuel and Blanket Air Systems Flame Ionization Detection CellIntroduction Anode Power Supply Electrometer-AmplifierFlame Guard Circuit Flame Ignition CircuitUnpacking the Analyzer Mounting the AnalyzerElectrical Power Connections User ConnectionsGas Connections Installation Electrical Connections Equipment Interface Connector Pin Arrangement Pin Function Threshold Alarm Pin Contact Pin out of 50 pins D-Sub Connector Pin Out TableCommands via RS-232 Input 4 RS-232 PortRequired RS-232 Options Supporting GasesSample Bypass Vent Placing the System in Operation Activating the Support Gases1 Air Carrier GasFlame Ignition FuelAnalyzer Operation Verification of the Flame Guard CircuitIgnition and/or Flame Guard Circuit Failure Sample PumpOperation Default Parameters Style ConventionsNavigation and Data Entry EnterEscape Menu Structure LineMain Menus Autocal is not available for standard configurations Logout Span ANALOG-OUT Adjust Setting up an AUTO-CAL Adjusting the GainPassword Protection Entering a Password Repeat Password Entry Screen Logging Out System Self-Diagnostic Test Model ScreenSpan Alarms Function AL-2 Source SMP Range Function Change Stream is not available for standard configurations Changing StreamAnalog output Adjustment Advanced User Functions StandbyTiming Recommended Spare Parts List Qty DescriptionDrawing List Analytical Instruments