Installation
3.3.3.5ALARM RELAYS
The nine
Threshold Alarm 1:
Can be configured as high (actuates when concentration is above threshold), or low (actuates when concentration is below threshold).
Can be configured as failsafe or
Threshold Alarm 2:
Can be configured as high (actuates when concentration is above threshold), or low (actuates when concentration is below threshold).
System Alarm:
Actuates when power is removed from analyzer. It is triggered if 4030 fails to ignite after five times.
The System Alarm may be reset by selecting the STANDBY function in the menu to send to sleep mode, then press the ENTER or ESCAPE key to restart the analyzer. However, if the condition that caused the 4030 to fail ignition is not remedied, the 4030 will likely fail to ignite again, and the System Alarm will be
Further detail on flame ignition failure can be found in Chapter 4, Section 4.3.1.
Teledyne Analytical Instruments | 21 |