Page 48

Operation

________________________________________________

<Enter or Escape> Start the diagnostic testing routine by pressing <Enter> or cancel out by pressing <Escape>.

If you pressed <Enter> the self-test routine will begin and after a few moments the results will appear onscreen. The module is functioning properly if it is followed by OK otherwise it is followed by BAD. In this case, please contact TAI Customer Service for an explanation of the problem.

To return the analyzer to the MAIN MENU, press <Enter> after the results screen.

If you pressed <Escape> you will be returned to the Analyze Mode.

4.4.10 Span

The SPAN function is used to calibrate the analyzer.

The analyzer is calibrated using span gas as described in Section 4.1. This section assumes that this gas has been properly connected and the line checked for leaks.

To initiate a pan calibration:

<▲▼> From the MAIN MENU, scroll down to the SPAN function.

<Enter> Press <Enter> to activate the SPAN function.

________________________________________________

-> THC

Span?

YES

SMP

Span?

NO

THC

Sp:

8.00

ppm

SMP

Sp:

0.00

ppm

Span Begin

________________________________________________

Teledyne Analytical Instruments

46

Image 48
Contents Series 4030 Heated THC Copyright 2004 Teledyne Analytical Instruments Safety Messages Series Additional Safety Information Series Procedure for Removal of Internal Inaccessible Shock Hazards Series Main Features of the Analyzer IntroductionPrinciple of Operation Analyzer DescriptionApplications IntroductionSample System Input PortingDetection Cell Gas Flow Control SystemFuel and Blanket Air Systems Flame Ionization Detection CellIntroduction Anode Power Supply Electrometer-AmplifierFlame Guard Circuit Flame Ignition CircuitUnpacking the Analyzer Mounting the AnalyzerElectrical Power Connections User ConnectionsGas Connections Installation Electrical Connections Equipment Interface Connector Pin Arrangement Pin Function Threshold Alarm Pin Contact Pin out of 50 pins D-Sub Connector Pin Out TableCommands via RS-232 Input 4 RS-232 PortRequired RS-232 Options Supporting GasesSample Bypass Vent Placing the System in Operation Activating the Support Gases1 Air Carrier GasFlame Ignition FuelAnalyzer Operation Verification of the Flame Guard CircuitIgnition and/or Flame Guard Circuit Failure Sample PumpOperation Default Parameters Style ConventionsNavigation and Data Entry EnterEscape Menu Structure LineMain Menus Autocal is not available for standard configurations Logout Span ANALOG-OUT Adjust Setting up an AUTO-CAL Adjusting the GainPassword Protection Entering a Password Repeat Password Entry Screen Logging Out System Self-Diagnostic Test Model ScreenSpan Alarms Function AL-2 Source SMP Range Function Change Stream is not available for standard configurations Changing StreamAnalog output Adjustment Advanced User Functions StandbyTiming Recommended Spare Parts List Qty DescriptionDrawing List Analytical Instruments