Teledyne HEATED THC, SERIES 4030 manual Sample Bypass Vent

Page 28

Installation

CAUTION: THE GASES USED MUST BE OF THE HIGHEST QUALITY, ULTRA ZERO GRADES, AS SHOWN BELOW. FAILURE TO DO SO WILL RESULT IN CONTAMINATION AND FAILURE TO DETECT LOW CONCENTRATIONS OF HYDROCARBONS.

AIR: USE ULTRA ZERO GRADE AIR WITH THC LESS THAN

0.1PPM. DO NOT USE OIL PUMPED AIR UNDER ANY CIRCUMSTANCES.

FUEL: USE ULTRA HIGH PURITY (UHP) GRADE HYDROGEN WITH THC LESS THAN 0.5 PPM.

CARRIER GAS: USE ULTRA HIGH PURITY GRADE NITROGEN GAS, WITH THC LESS THAN 0.1 PPM.

3.3.5.2EFFLUENT

All the gases introduced into the detection cell vent from one fitting at the rear of the analyzer. TAI recommends that the cell be permitted to vent directly to the atmosphere wherever possible.

If a vent line is required, the installation must include a drop-out pot to collect the water that is formed by the burning of the hydrogen fuel. The vent line must be constructed so that water and dirt cannot collect in it.

3.3.5.3SAMPLE BYPASS VENT

The sample bypassed by the back-pressure regulation system vents from a separate port at the rear of the analyzer. If a vent line is required, it must be installed so that water and dirt cannot accumulate in it.

3.3.5.4FUEL AND AIR CONNECTIONS

The fuel used to provide combustion should be Hydrogen gas, zero quality with certified THC of less than 0.5 ppm and a pressure of 30 psig. The compressed air, ultra zero gas quality with THC less than 0.1 ppm should be used at a controlled pressure of 30 psig Connect the fuel and air sources to the instrument according to the gas connection diagram included at the back of this manual.

Teledyne Analytical Instruments

26

Image 28
Contents Series 4030 Heated THC Copyright 2004 Teledyne Analytical Instruments Safety Messages Series Additional Safety Information Series Procedure for Removal of Internal Inaccessible Shock Hazards Series Main Features of the Analyzer IntroductionPrinciple of Operation Analyzer DescriptionApplications IntroductionSample System Input PortingDetection Cell Gas Flow Control SystemFuel and Blanket Air Systems Flame Ionization Detection CellIntroduction Anode Power Supply Electrometer-AmplifierFlame Guard Circuit Flame Ignition CircuitUnpacking the Analyzer Mounting the AnalyzerUser Connections Electrical Power ConnectionsGas Connections Installation Electrical Connections Equipment Interface Connector Pin Arrangement Pin Function Threshold Alarm Pin Contact Pin out of 50 pins D-Sub Connector Pin Out TableCommands via RS-232 Input 4 RS-232 PortRequired RS-232 Options Supporting GasesSample Bypass Vent Placing the System in Operation Activating the Support Gases1 Air Carrier GasFlame Ignition FuelAnalyzer Operation Verification of the Flame Guard CircuitIgnition and/or Flame Guard Circuit Failure Sample PumpOperation Default Parameters Style ConventionsNavigation and Data Entry EnterEscape Menu Structure LineMain Menus Autocal is not available for standard configurations Logout Span ANALOG-OUT Adjust Setting up an AUTO-CAL Adjusting the GainPassword Protection Entering a Password Repeat Password Entry Screen Logging Out System Self-Diagnostic Test Model ScreenSpan Alarms Function AL-2 Source SMP Range Function Change Stream is not available for standard configurations Changing StreamAnalog output Adjustment Standby Advanced User FunctionsTiming Recommended Spare Parts List Qty DescriptionDrawing List Analytical Instruments