Introduction

nominally 125 volts. Output tolerance is ±10 volts from the specified 125 volts.

2.3.3 Flame Guard Circuit

A thermistor-controlled, transistorized switching circuit is employed to operate a relay in the event of a flameout condition. A panel indicator light and fuel shut-off solenoid valve are operated by the relay to alarm personnel that a flameout condition has occurred. The fuel shut-off solenoid valve stops the hydrogen flow.

2.3.4 Flame Ignition Circuit

The flame ignition circuit includes the anode-igniter electrode (in the detector cell), a transformer, and processor-controlled relay. The circuit is automatically energized when the FID cools due to lack of flame.

If automatic ignition fails five times, there will be a message that reports this, and the flame can be manually ignited by pressing simultaneously the Up and Down key.

Teledyne Analytical Instruments

14

Image 16
Contents Series 4030 Heated THC Copyright 2004 Teledyne Analytical Instruments Safety Messages Series Additional Safety Information Series Procedure for Removal of Internal Inaccessible Shock Hazards Series Main Features of the Analyzer IntroductionPrinciple of Operation Analyzer DescriptionApplications IntroductionSample System Input PortingDetection Cell Gas Flow Control SystemFuel and Blanket Air Systems Flame Ionization Detection CellIntroduction Anode Power Supply Electrometer-AmplifierFlame Guard Circuit Flame Ignition CircuitUnpacking the Analyzer Mounting the AnalyzerUser Connections Electrical Power ConnectionsGas Connections Installation Electrical Connections Equipment Interface Connector Pin Arrangement Pin Function Threshold Alarm Pin Contact Pin out of 50 pins D-Sub Connector Pin Out TableCommands via RS-232 Input 4 RS-232 PortRequired RS-232 Options Supporting GasesSample Bypass Vent Placing the System in Operation Activating the Support Gases1 Air Carrier GasFlame Ignition FuelAnalyzer Operation Verification of the Flame Guard CircuitIgnition and/or Flame Guard Circuit Failure Sample PumpOperation Default Parameters Style ConventionsNavigation and Data Entry EnterEscape Menu Structure LineMain Menus Autocal is not available for standard configurations Logout Span ANALOG-OUT Adjust Setting up an AUTO-CAL Adjusting the GainPassword Protection Entering a Password Repeat Password Entry Screen Logging Out System Self-Diagnostic Test Model ScreenSpan Alarms Function AL-2 Source SMP Range Function Change Stream is not available for standard configurations Changing StreamAnalog output Adjustment Standby Advanced User FunctionsTiming Recommended Spare Parts List Qty DescriptionDrawing List Analytical Instruments