Teledyne HEATED THC, SERIES 4030 manual Pin Function

Page 22

Installation

Both the first and third set of outputs as well as the second and fourth set of outputs may be set to any of the following functions:

-THC -- Total hydrocarbon as measured in the first phase of detection. The output is the percentage of the selected range.

-SMP -- An alternate measurement that occurs during the second phase of detection. The output represents the percentage of the selected range.

-WAVE – Analogous to the continuous output of the FID.

-RID – Indicates the selected range. Low Range is at 1/5 of maximum output. Mid Range is at 1/2 of maximum output. High Range is 4/5 of maximum output.

Table 3-1: Analog Output Connections

Pin Function

3Channel 2, + 4-20 mA, floating

4Channel 2, – 4-20 mA, floating

5Channel 1, + 4-20 mA, floating

6Channel 1, – 4-20 mA, floating

8Channel 2, 0-1 VDC

23Channel 2, 0-1 VDC

24Channel 1, 0-1 VDC

7Channel 1, negative ground

Teledyne Analytical Instruments

20

Image 22
Contents Series 4030 Heated THC Copyright 2004 Teledyne Analytical Instruments Safety Messages Series Additional Safety Information Series Procedure for Removal of Internal Inaccessible Shock Hazards Series Main Features of the Analyzer IntroductionPrinciple of Operation Analyzer DescriptionApplications IntroductionSample System Input PortingFuel and Blanket Air Systems Detection CellGas Flow Control System Flame Ionization Detection CellIntroduction Anode Power Supply Electrometer-AmplifierFlame Guard Circuit Flame Ignition CircuitUnpacking the Analyzer Mounting the AnalyzerUser Connections Electrical Power ConnectionsGas Connections Installation Electrical Connections Equipment Interface Connector Pin Arrangement Pin Function Threshold Alarm Pin Contact Pin out of 50 pins D-Sub Connector Pin Out TableCommands via RS-232 Input 4 RS-232 PortRequired RS-232 Options Supporting GasesSample Bypass Vent 1 Air Placing the System in OperationActivating the Support Gases Carrier GasFlame Ignition FuelIgnition and/or Flame Guard Circuit Failure Analyzer OperationVerification of the Flame Guard Circuit Sample PumpOperation Default Parameters Style ConventionsNavigation and Data Entry EnterEscape Menu Structure LineMain Menus Autocal is not available for standard configurations Logout Span ANALOG-OUT Adjust Setting up an AUTO-CAL Adjusting the GainPassword Protection Entering a Password Repeat Password Entry Screen Logging Out System Self-Diagnostic Test Model ScreenSpan Alarms Function AL-2 Source SMP Range Function Change Stream is not available for standard configurations Changing StreamAnalog output Adjustment Standby Advanced User FunctionsTiming Recommended Spare Parts List Qty DescriptionDrawing List Analytical Instruments