Teledyne 3300PA operating instructions Electrochemical Reactions, Effect of Pressure

Page 13

Percent Oxygen Analyzer

Operational Theory 2

 

 

 

 

 

 

 

 

 

The space between the active elements is filled by a structure saturated with electrolyte. Cathode and anode are wet by this common pool. They each have a conductor connecting them, through some electrical circuitry, to one of the external contacts in the connector receptacle, which is on the top of the cell.

2.2.3 Electrochemical Reactions

The sample gas diffuses through the Teflon membrane. Any oxygen in the sample gas is reduced on the surface of the cathode by the following HALF REACTION:

O

2

+ 2H O + 4e4OH

(cathode)

 

2

 

(Four electrons combine with one oxygen molecule—in the presence of water from the electrolyte—to produce four hydroxyl ions.)

When the oxygen is reduced at the cathode, lead is simultaneously oxidized at the anode by the following HALF REACTION:

2(Pb + 2OH)2(Pb+2 + H O) + 4e

(anode)

2

 

(Two electrons are transferred for each atom of lead that is oxidized. TWO ANODE REACTIONS balance one cathode reaction to transfer four elec- trons.)

The electrons released at the surface of the anode flow to the cathode surface when an external electrical path is provided. The current is proportional to the amount of oxygen reaching the cathode. It is measured and used to determine the oxygen concentration in the gas mixture.

The overall reaction for the fuel cell is the SUM of the half reactions above,

or:

2Pb + O2 2PbO

(These reactions will hold as long as no gaseous components capable of oxidizing lead are present in the sample. The only likely components are the halogens—iodine,bromine,chlorineandfluorine.)

The output of the fuel cell is limited by (1) the amount of oxygen in the cell at the time and (2) the amount of stored anode material.

In the absence of oxygen, no current is generated.

2.2.4 The Effect of Pressure

In order to state the amount of oxygen present in the sample as a percent- age of the gas mixture, it is necessary that the sample diffuse into the cell under constant pressure.

Teledyne Analytical Instruments

2-3

Image 13
Contents Model 3300PA Copyright 1999 Teledyne Analytical Instruments Contents Appendix MaintenanceCombustible GAS Usage Warning This page left intentionally blank Introduction Overview Main Features of the AnalyzerPercent Oxygen Analyzer Introduction Introduction Model 3300PA Front Panel DescriptionRear Panel Rear Panel DescriptionAnalog Outputs Micro-Fuel Cell Sensor Operational Theory IntroductionPercent Oxygen Analyzer Operational Theory Principles of OperationAnatomy of a Micro-Fuel Cell Operational Theory Model 3300PAEffect of Pressure Electrochemical ReactionsCharacteristic Input/Output Curve for a Micro-Fuel Cell Calibration CharacteristicsElectronics Signal ProcessingGeneral Operational Theory Model 3300PA Percent Oxygen Analyzer Installation InstallationUnpacking the Analyzer Control Unit Installation Installation Model 3300PAInstalling the Micro-Fuel Cell Location and MountingPercent Oxygen Analyzer Installation Sensor Fail Gas ConnectionsVacuum Service Option Installation Model 3300 PA Installation ChecklistPercent Oxygen Analyzer Operation Operation IntroductionSetting the Analysis Ranges Using the Function and Data Entry ButtonsOperation Model 3300PA Set Alarm Setting the Alarm SetpointsPercent Oxygen Analyzer Operation HI Range LO RangeSensor Fail Alarm Calibration Selecting a Fixed Range or AutorangingSupplementary Information Percent Oxygen Analyzer Maintenance MaintenanceRemove Power to Unit before replacing the fuse Replacing the FuseMaintenance Model 3300PA Sensor Installation or ReplacementWhen to Replace a Sensor Ordering and Handling of Spare SensorsRemoving the Micro-Fuel Cell Installing a Micro-Fuel CellCell Warranty Conditions Percent Oxygen Analyzer Appendix Appendix SpecificationsAppendix Model 3300PA Spare Parts ListMiscellaneous Reference DrawingAppendix Model 3300PA