Teledyne 3300PA operating instructions Removing the Micro-Fuel Cell, Installing a Micro-Fuel Cell

Page 31

Percent Oxygen Analyzer

Maintenance 5

 

 

 

 

 

 

 

 

 

ted to enter, cell-life will be compromised.

WARNING: The sensor used in the Model 3300PA uses electrolytes which contain substances that are extremely harmful if touched, swallowed, or inhaled. Avoid contact with ANY

fluid or powder in or around the unit. What may appear to be

plain water could contain one of these toxic substances. In case of eye contact, immediately flush eyes with water for at least 15 minutes. Call physician. (See Appendix, Material Safety Data Sheet— MSDS).

5.2.3 Removing the Micro-Fuel Cell

To remove a spent or damaged Micro-Fuel Cell:

1.Disconnect the Power Source from the Unit.

2.Open Sensor Housing door on the front panel.

3.Unplug the connector from the spent cell.

4.Remove the spent cell by unscrewing it, counterclockwise, from the cell block.

5.Dispose of the cell in a safe manner, and in accordance with local laws.

5.2.4 Installing a Micro-Fuel Cell

To install a new Micro-Fuel Cell:

Teledyne Analytical Instruments

5-3

Image 31
Contents Model 3300PA Copyright 1999 Teledyne Analytical Instruments Contents Appendix MaintenanceCombustible GAS Usage Warning This page left intentionally blank Introduction Overview Main Features of the AnalyzerPercent Oxygen Analyzer Introduction Introduction Model 3300PA Front Panel DescriptionRear Panel Rear Panel DescriptionAnalog Outputs Principles of Operation Operational Theory IntroductionMicro-Fuel Cell Sensor Percent Oxygen Analyzer Operational TheoryAnatomy of a Micro-Fuel Cell Operational Theory Model 3300PAEffect of Pressure Electrochemical ReactionsCharacteristic Input/Output Curve for a Micro-Fuel Cell Calibration CharacteristicsElectronics Signal ProcessingGeneral Operational Theory Model 3300PA Percent Oxygen Analyzer Installation InstallationUnpacking the Analyzer Location and Mounting Installation Model 3300PAControl Unit Installation Installing the Micro-Fuel CellPercent Oxygen Analyzer Installation Sensor Fail Gas ConnectionsVacuum Service Option Installation Model 3300 PA Installation ChecklistPercent Oxygen Analyzer Operation Operation IntroductionSetting the Analysis Ranges Using the Function and Data Entry ButtonsOperation Model 3300PA LO Range Setting the Alarm SetpointsSet Alarm Percent Oxygen Analyzer Operation HI RangeSensor Fail Alarm Calibration Selecting a Fixed Range or AutorangingSupplementary Information Replacing the Fuse MaintenancePercent Oxygen Analyzer Maintenance Remove Power to Unit before replacing the fuseOrdering and Handling of Spare Sensors Sensor Installation or ReplacementMaintenance Model 3300PA When to Replace a SensorRemoving the Micro-Fuel Cell Installing a Micro-Fuel CellCell Warranty Conditions Percent Oxygen Analyzer Appendix Appendix SpecificationsAppendix Model 3300PA Spare Parts ListMiscellaneous Reference DrawingAppendix Model 3300PA