Teledyne 3300PA Setting the Alarm Setpoints, Percent Oxygen Analyzer Operation HI Range, LO Range

Page 25

Percent Oxygen Analyzer

Operation 4

 

 

 

 

 

 

 

 

 

4.3.1 HI Range

Setting the HI Range fullscale value defines the LEAST sensitive analysis range to be used. To set the HI Range:

1.

Press the SET HI RANGE Function button once.

2.

Immediately (within 5 seconds) press either the or button to

 

raise or lower the displayed value, as required, until the display

 

reads the desired fullscale percent concentration.

4.3.2 LO Range

Setting the LO Range fullscale value defines the MOST sensitive range to be used. To set the LO Range:

1.

Press the SET LO RANGE Function button once.

2.

Immediately (within 5 seconds) press either the or button to

 

raise or lower the displayed value, as required, until the display

 

reads the desired fullscale percent concentration.

4.4Setting the Alarm Setpoints

The alarm setpoints can be adjusted over the full range of the analyzer (0-25% oxygen content). They are set as a percent of oxygen content, so that an alarm set to indicate 9.6 on the display will activate at 9.6% O2 on any O2 range.

4.4.1 Set Alarm 1

Alarm 1 can be set either as a high or low alarm. To configure this alarm to your preferences:

1.Press the SET ALARM 1 function button once.

2.The display will flash either HI or LO depending on what the alarm was configured in the last time. If the alarm configuration must be changed, use the Up or Down keys to toggle the alarm from HI to LO or viceversa (if within 5 seconds no key is pressed, the instrument will return to the sample mode and display oxygen concentration).

3.To change the setting at which the alarm will be actuated, press the SET ALARM 1 function button once more. The alarm setpoint will flash on the LED display. Press either the Up or Down keys to raise or lower the displayed value, as required,

Teledyne Analytical Instruments

4-3

Image 25
Contents Model 3300PA Copyright 1999 Teledyne Analytical Instruments Contents Appendix MaintenanceCombustible GAS Usage Warning This page left intentionally blank Introduction Overview Main Features of the AnalyzerPercent Oxygen Analyzer Introduction Introduction Model 3300PA Front Panel DescriptionRear Panel Rear Panel DescriptionAnalog Outputs Micro-Fuel Cell Sensor Operational Theory IntroductionPercent Oxygen Analyzer Operational Theory Principles of OperationAnatomy of a Micro-Fuel Cell Operational Theory Model 3300PAEffect of Pressure Electrochemical ReactionsCharacteristic Input/Output Curve for a Micro-Fuel Cell Calibration CharacteristicsElectronics Signal ProcessingGeneral Operational Theory Model 3300PA Percent Oxygen Analyzer Installation InstallationUnpacking the Analyzer Control Unit Installation Installation Model 3300PAInstalling the Micro-Fuel Cell Location and MountingPercent Oxygen Analyzer Installation Sensor Fail Gas ConnectionsVacuum Service Option Installation Model 3300 PA Installation ChecklistPercent Oxygen Analyzer Operation Operation IntroductionSetting the Analysis Ranges Using the Function and Data Entry ButtonsOperation Model 3300PA Set Alarm Setting the Alarm SetpointsPercent Oxygen Analyzer Operation HI Range LO RangeSensor Fail Alarm Calibration Selecting a Fixed Range or AutorangingSupplementary Information Percent Oxygen Analyzer Maintenance MaintenanceRemove Power to Unit before replacing the fuse Replacing the FuseMaintenance Model 3300PA Sensor Installation or ReplacementWhen to Replace a Sensor Ordering and Handling of Spare SensorsRemoving the Micro-Fuel Cell Installing a Micro-Fuel CellCell Warranty Conditions Percent Oxygen Analyzer Appendix Appendix SpecificationsAppendix Model 3300PA Spare Parts ListMiscellaneous Reference DrawingAppendix Model 3300PA