Teledyne 3300PA operating instructions Sensor Fail Alarm

Page 26

4 Operation

Model 3300PA

 

 

 

 

 

 

 

 

 

until the display reads the desired percent concentration. (if within 5 seconds no key is pressed, the instrument will return to the sample mode and display oxygen concentration).

After setting the value wait for the unit to time out of this mode (ap- proximately 5 seconds) and return to displaying oxygen concentration.

4.4.2 Set Alarm 2

Alarm 2 can be set either as a high or low alarm. To configure this alarm to your preferences:

1.Press the SET ALARM 2 function button once.

2.The display will flash either HI or LO depending on what the alarm was configured in the last time. If the alarm configuration must be changed, use the Up or Down keys to toggle the alarm from HI to LO or viceversa (if within 5 seconds no key is pressed, the instrument will return to the sample mode and display oxygen concentration).

3.To change the setting at which the alarm will be actuated, press the SET ALARM 2 function button once more. The alarm setpoint will flash on the LED display. Press either the Up or Down keys to raise or lower the displayed value, as required, until the display reads the desired percent concentration. (if within 5 seconds no key is pressed, the instrument will return to the sample mode and display oxygen concentration)

After setting the value wait for the unit to time out of this mode (ap- proximately 5 seconds) and return to displaying oxygen concentration.

4.4.3 Sensor Fail Alarm

The SENSOR FAIL alarm is factory set to a reading less than 0.05% O2. Should this alarm trigger the ALARM Indicator below the SET Func- tion buttons will blink, and the alarm relay contact dedicated to this function will change state.

4-4

Teledyne Analytical Instruments

Image 26
Contents Model 3300PA Copyright 1999 Teledyne Analytical Instruments Contents Maintenance AppendixCombustible GAS Usage Warning This page left intentionally blank Percent Oxygen Analyzer Introduction Main Features of the AnalyzerIntroduction Overview Front Panel Description Introduction Model 3300PARear Panel Description Rear PanelAnalog Outputs Percent Oxygen Analyzer Operational Theory Operational Theory IntroductionMicro-Fuel Cell Sensor Principles of OperationOperational Theory Model 3300PA Anatomy of a Micro-Fuel CellElectrochemical Reactions Effect of PressureCalibration Characteristics Characteristic Input/Output Curve for a Micro-Fuel CellGeneral Signal ProcessingElectronics Operational Theory Model 3300PA Unpacking the Analyzer InstallationPercent Oxygen Analyzer Installation Installing the Micro-Fuel Cell Installation Model 3300PAControl Unit Installation Location and MountingPercent Oxygen Analyzer Installation Gas Connections Sensor FailVacuum Service Option Installation Checklist Installation Model 3300 PAOperation Introduction Percent Oxygen Analyzer OperationOperation Model 3300PA Using the Function and Data Entry ButtonsSetting the Analysis Ranges Percent Oxygen Analyzer Operation HI Range Setting the Alarm SetpointsSet Alarm LO RangeSensor Fail Alarm Selecting a Fixed Range or Autoranging CalibrationSupplementary Information Remove Power to Unit before replacing the fuse MaintenancePercent Oxygen Analyzer Maintenance Replacing the FuseWhen to Replace a Sensor Sensor Installation or ReplacementMaintenance Model 3300PA Ordering and Handling of Spare SensorsInstalling a Micro-Fuel Cell Removing the Micro-Fuel CellCell Warranty Conditions Appendix Specifications Percent Oxygen Analyzer AppendixSpare Parts List Appendix Model 3300PAReference Drawing MiscellaneousAppendix Model 3300PA