Teledyne 3300PA operating instructions Replacing the Fuse, Percent Oxygen Analyzer Maintenance

Page 29

Percent Oxygen Analyzer

Maintenance 5

 

 

 

 

 

 

 

 

 

Overvoltage

Category II

Maintenance

Aside from normal cleaning and checking for leaks at the gas connections, the Model 3300PA should not require any maintenance beyond replacement of expended Micro-Fuel Cells, and perhaps a blown fuse. Routine maintenance includes occasional recalibration, as described in chapter 4, Operation.

5.1Replacing the Fuse

Remove Power to Unit before replacing the fuse.

When a fuse blows, check first to determine the cause, then replace the fuse using the following procedure:

1.Disconnect the AC power and place the power switch located on the rear panel in the O position. Remove the power cord from the receptacle.

2.The fuse receptacle is located in the power cord receptacle assembly in the upper left-hand corner of the rear panel. See Figure 5-1.

Fuse

Figure 5-1: AC Fuse Replacement

3.Insert a small flat-blade screwdriver into the slot in the receptacle wall nearest the fuse and gently pry open the fuse receptacle. The

Teledyne Analytical Instruments

5-1

Image 29
Contents Model 3300PA Copyright 1999 Teledyne Analytical Instruments Contents Appendix MaintenanceCombustible GAS Usage Warning This page left intentionally blank Percent Oxygen Analyzer Introduction Main Features of the AnalyzerIntroduction Overview Introduction Model 3300PA Front Panel DescriptionRear Panel Rear Panel DescriptionAnalog Outputs Micro-Fuel Cell Sensor Operational Theory IntroductionPercent Oxygen Analyzer Operational Theory Principles of OperationAnatomy of a Micro-Fuel Cell Operational Theory Model 3300PAEffect of Pressure Electrochemical ReactionsCharacteristic Input/Output Curve for a Micro-Fuel Cell Calibration CharacteristicsGeneral Signal ProcessingElectronics Operational Theory Model 3300PA Unpacking the Analyzer InstallationPercent Oxygen Analyzer Installation Control Unit Installation Installation Model 3300PAInstalling the Micro-Fuel Cell Location and MountingPercent Oxygen Analyzer Installation Sensor Fail Gas ConnectionsVacuum Service Option Installation Model 3300 PA Installation ChecklistPercent Oxygen Analyzer Operation Operation IntroductionOperation Model 3300PA Using the Function and Data Entry ButtonsSetting the Analysis Ranges Set Alarm Setting the Alarm SetpointsPercent Oxygen Analyzer Operation HI Range LO RangeSensor Fail Alarm Calibration Selecting a Fixed Range or AutorangingSupplementary Information Percent Oxygen Analyzer Maintenance MaintenanceRemove Power to Unit before replacing the fuse Replacing the FuseMaintenance Model 3300PA Sensor Installation or ReplacementWhen to Replace a Sensor Ordering and Handling of Spare SensorsRemoving the Micro-Fuel Cell Installing a Micro-Fuel CellCell Warranty Conditions Percent Oxygen Analyzer Appendix Appendix SpecificationsAppendix Model 3300PA Spare Parts ListMiscellaneous Reference DrawingAppendix Model 3300PA