Teledyne 3300PA operating instructions Appendix Specifications, Percent Oxygen Analyzer Appendix

Page 33

Percent Oxygen Analyzer

Appendix

 

 

 

 

 

 

 

 

 

Appendix

A.1 Specifications

Ranges:

0-3 % and 0-10 % oxygen (Standard

 

Ranges), and 0-25 % (nominal) Calibration

 

Range. User selectable % Ranges can be set

 

between 1% and 25 % oxygen (optional

 

100% range).

SignalOutput:

Voltage: 0–10 V dc, negative ground

 

Current: 4-20 mA, negative ground

Range ID:

0-10 V dc.

Display:

Lightemittingdiodedisplay.

Alarms:

Two customer selectable high or low fully

 

adjustable alarms. One cell failure alarm.

 

Alarm relays form "C" contacts, dry contacts

 

rated at 2A for 30VDC, 0.5A for 115VAC.

System Operating Temp:

0-50 °C

Accuracy:

±2% of full scale at constant temperature

 

±5% of full scale through operating temp.

 

range once temp. equilibrium is reached. (At

 

3 % and higher user defined ranges.)

Response Time:

90% in less than 8 seconds at 25 °C.

System Power Requirements:

AC (100-240 VAC, 50/60Hz), 10W.

System Enclosure:

Rack mount with handles. Dimensions

 

(Approx.) 19 W x 5 H x 5 D.

Sensor Type:

Micro Fuel Cell Class I-22.

Teledyne Analytical Instruments

A-1

Image 33
Contents Model 3300PA Copyright 1999 Teledyne Analytical Instruments Contents Appendix MaintenanceCombustible GAS Usage Warning This page left intentionally blank Main Features of the Analyzer Introduction OverviewPercent Oxygen Analyzer Introduction Introduction Model 3300PA Front Panel DescriptionRear Panel Rear Panel DescriptionAnalog Outputs Micro-Fuel Cell Sensor Operational Theory IntroductionPercent Oxygen Analyzer Operational Theory Principles of OperationAnatomy of a Micro-Fuel Cell Operational Theory Model 3300PAEffect of Pressure Electrochemical ReactionsCharacteristic Input/Output Curve for a Micro-Fuel Cell Calibration CharacteristicsSignal Processing ElectronicsGeneral Operational Theory Model 3300PA Installation Percent Oxygen Analyzer InstallationUnpacking the Analyzer Control Unit Installation Installation Model 3300PAInstalling the Micro-Fuel Cell Location and MountingPercent Oxygen Analyzer Installation Sensor Fail Gas ConnectionsVacuum Service Option Installation Model 3300 PA Installation ChecklistPercent Oxygen Analyzer Operation Operation IntroductionUsing the Function and Data Entry Buttons Setting the Analysis RangesOperation Model 3300PA Set Alarm Setting the Alarm SetpointsPercent Oxygen Analyzer Operation HI Range LO RangeSensor Fail Alarm Calibration Selecting a Fixed Range or AutorangingSupplementary Information Percent Oxygen Analyzer Maintenance MaintenanceRemove Power to Unit before replacing the fuse Replacing the FuseMaintenance Model 3300PA Sensor Installation or ReplacementWhen to Replace a Sensor Ordering and Handling of Spare SensorsRemoving the Micro-Fuel Cell Installing a Micro-Fuel CellCell Warranty Conditions Percent Oxygen Analyzer Appendix Appendix SpecificationsAppendix Model 3300PA Spare Parts ListMiscellaneous Reference DrawingAppendix Model 3300PA