Teledyne operating instructions Operational Theory Model 3300PA

Page 16

2 Operational Theory

Model 3300PA

 

 

 

 

 

 

 

 

 

In the presence of oxygen the cell generates a current. This current flows through an internal thermistor compensation network.

The output of the sensor is converted to voltage in the millivolt range. This output is fed to a voltage amplifier. The internal thermistor network provides temperature compensation of the sensor output. The resistance of the network changes with temperature, compensating for the changes of the microfuel cell output to temperature.

The output from the sensor amplifier is sent to an analog to digital converter (ADC), and the resulting digital concentration signal is sent to the microcontroller.

The digital concentration signal along with input from the front panel buttons (KEYBOARD) is processed by the microcontroller, and appropriate output signals are directed to the display and alarm relays. The same digital information is also sent to a 12-bit digital to analog converter (DAC) that produces the 0-10

Vdc analog concentration signal and the 0-10 V dc analog range ID output. A current to voltage converter (E–I CONV) produces the 4-20 mA dc analog concentrationsignal.

2-6

Teledyne Analytical Instruments

Image 16
Contents Model 3300PA Copyright 1999 Teledyne Analytical Instruments Contents Maintenance AppendixCombustible GAS Usage Warning This page left intentionally blank Introduction Overview Main Features of the AnalyzerPercent Oxygen Analyzer Introduction Front Panel Description Introduction Model 3300PARear Panel Description Rear PanelAnalog Outputs Operational Theory Introduction Micro-Fuel Cell SensorPercent Oxygen Analyzer Operational Theory Principles of OperationOperational Theory Model 3300PA Anatomy of a Micro-Fuel CellElectrochemical Reactions Effect of PressureCalibration Characteristics Characteristic Input/Output Curve for a Micro-Fuel CellElectronics Signal ProcessingGeneral Operational Theory Model 3300PA Percent Oxygen Analyzer Installation InstallationUnpacking the Analyzer Installation Model 3300PA Control Unit InstallationInstalling the Micro-Fuel Cell Location and MountingPercent Oxygen Analyzer Installation Gas Connections Sensor FailVacuum Service Option Installation Checklist Installation Model 3300 PAOperation Introduction Percent Oxygen Analyzer OperationSetting the Analysis Ranges Using the Function and Data Entry ButtonsOperation Model 3300PA Setting the Alarm Setpoints Set AlarmPercent Oxygen Analyzer Operation HI Range LO RangeSensor Fail Alarm Selecting a Fixed Range or Autoranging CalibrationSupplementary Information Maintenance Percent Oxygen Analyzer MaintenanceRemove Power to Unit before replacing the fuse Replacing the FuseSensor Installation or Replacement Maintenance Model 3300PAWhen to Replace a Sensor Ordering and Handling of Spare SensorsInstalling a Micro-Fuel Cell Removing the Micro-Fuel CellCell Warranty Conditions Appendix Specifications Percent Oxygen Analyzer AppendixSpare Parts List Appendix Model 3300PAReference Drawing MiscellaneousAppendix Model 3300PA