Percent Oxygen Analyzer | Installation 3 | |
|
|
|
|
|
|
|
|
|
atmospheric pressure. Restricting the sample vent line will result in pressur- izing the sensor and altering the O2 reading.
The pressure required will depend on the sampling system. When venting into a constant pressure, such as the atmosphere, controlling input pressure is simple. If you are venting into a system of varying pressure, then some form of pressure regulation is required.
Figure 3-3: Piping Diagram
(standard piping diagram for positive pressure service)
3.4.1 Vacuum Service Option
For vacuum service plumbing reference figure
Adjust the flow rate on a vacuum service system as follows:
Select Sample with the
Teledyne Analytical Instruments |