Teledyne 3300PA operating instructions Calibration Characteristics

Page 14

2 Operational Theory

Model 3300PA

 

 

 

 

 

 

 

 

 

If the pressure changes, the rate that oxygen reaches the cathode through the diffusing membrane will also increase. The electron transfer, and therefore the external current, will increase, even though the proportion of oxygen has not changed.

Fortunately, Dalton's Law confirms that every gas in a mixture contributes the same pressure to the mixture that it would exert if it were alone in the same amount in that same volume. This means that as long as the total pressure of the sample remains constant, the mixture can change, but the diffusion of the oxygen will be affected only by the concentration of the oxygen.

For this reason, the sample system supplying sample gas to the cell should be designed to keep the pressure on the diffusion membrane constant.

2.2.5 Calibration Characteristics

Given that the total pressure of the sample gas at the surface of the Micro- Fuel Cell input is constant, a convenient characteristic of the cell is that the current produced in an external circuit of constant impedance is directly propor- tional to the rate at which oxygen molecules reach the cathode, and this rate is directly proportional to the concentration of oxygen in the gaseous mixture. In other words it has a linear characteristic curve, as shown in Figure 2-2. Measur- ing circuits do not have to compensate for nonlinearities.

Figure 2-2. Characteristic Input/Output Curve for a Micro-Fuel Cell

In addition, since there is zero output in the absence oxygen, the character- istic curve has an absolute zero. The cell itself does not need to be zeroed.

2-4

Teledyne Analytical Instruments

Image 14
Contents Model 3300PA Copyright 1999 Teledyne Analytical Instruments Contents Maintenance AppendixCombustible GAS Usage Warning This page left intentionally blank Percent Oxygen Analyzer Introduction Main Features of the AnalyzerIntroduction Overview Front Panel Description Introduction Model 3300PARear Panel Description Rear PanelAnalog Outputs Percent Oxygen Analyzer Operational Theory Operational Theory IntroductionMicro-Fuel Cell Sensor Principles of OperationOperational Theory Model 3300PA Anatomy of a Micro-Fuel CellElectrochemical Reactions Effect of PressureCalibration Characteristics Characteristic Input/Output Curve for a Micro-Fuel CellGeneral Signal ProcessingElectronics Operational Theory Model 3300PA Unpacking the Analyzer InstallationPercent Oxygen Analyzer Installation Installing the Micro-Fuel Cell Installation Model 3300PAControl Unit Installation Location and MountingPercent Oxygen Analyzer Installation Gas Connections Sensor FailVacuum Service Option Installation Checklist Installation Model 3300 PAOperation Introduction Percent Oxygen Analyzer OperationOperation Model 3300PA Using the Function and Data Entry ButtonsSetting the Analysis Ranges Percent Oxygen Analyzer Operation HI Range Setting the Alarm SetpointsSet Alarm LO RangeSensor Fail Alarm Selecting a Fixed Range or Autoranging CalibrationSupplementary Information Remove Power to Unit before replacing the fuse MaintenancePercent Oxygen Analyzer Maintenance Replacing the FuseWhen to Replace a Sensor Sensor Installation or ReplacementMaintenance Model 3300PA Ordering and Handling of Spare SensorsInstalling a Micro-Fuel Cell Removing the Micro-Fuel CellCell Warranty Conditions Appendix Specifications Percent Oxygen Analyzer AppendixSpare Parts List Appendix Model 3300PAReference Drawing MiscellaneousAppendix Model 3300PA