Teledyne 3300PA Unpacking the Analyzer, Percent Oxygen Analyzer Installation

Page 17

Percent Oxygen Analyzer

Installation 3

 

 

 

 

 

 

 

 

 

Overvoltage Category II

Installation

Installation of the analyzer includes:

1.Unpacking the system.

2.Mounting the Control Unit to a 19" Rack.

3.Installing the Micro-Fuel Cell.

4.Making the electrical connections.

5.Making the gas connections.

6.Testing the installation.

CAUTIONS:

Read this chapter in its entirety before installing the units.

 

For indoor use only.

 

The Sample must be free of entrained solids or water. How-

 

ever, a high humidity sample is ideal, since it will prevent

 

water loss from the cell electrolyte.

 

The Micro-Fuel Cell sensor electrolyte is caustic. Do not

 

attempt to open it. Leaking or exhausted cells should be

 

disposed of in accordance with local regulations. Refer to the

 

Material Safety Data Sheet in the Appendix.

 

Any damage or scarring of the delicate permeable membrane

 

on the sensing end of the cell will require cell replacement.

 

Prevent contact with membrane by any solid object.

3.1

Unpacking the Analyzer

As soon as you receive the instrument, carefully unpack and inspect Control Unit, and any included accessories for damage. Immediately report any damage to the shipping agent. The analyzer is shipped with all the materials you need to install and prepare the system for operation.

CAUTION: Do not disturb the integrity of the cell package until the cell is to actually be used. If the cell package is punctured and air is permitted to enter, cell-life will be compromised.

Teledyne Analytical Instruments

3-1

Image 17
Contents Model 3300PA Copyright 1999 Teledyne Analytical Instruments Contents Appendix MaintenanceCombustible GAS Usage Warning This page left intentionally blank Percent Oxygen Analyzer Introduction Main Features of the AnalyzerIntroduction Overview Introduction Model 3300PA Front Panel DescriptionRear Panel Rear Panel DescriptionAnalog Outputs Micro-Fuel Cell Sensor Operational Theory IntroductionPercent Oxygen Analyzer Operational Theory Principles of OperationAnatomy of a Micro-Fuel Cell Operational Theory Model 3300PAEffect of Pressure Electrochemical ReactionsCharacteristic Input/Output Curve for a Micro-Fuel Cell Calibration CharacteristicsGeneral Signal ProcessingElectronics Operational Theory Model 3300PA Unpacking the Analyzer InstallationPercent Oxygen Analyzer Installation Control Unit Installation Installation Model 3300PAInstalling the Micro-Fuel Cell Location and MountingPercent Oxygen Analyzer Installation Sensor Fail Gas ConnectionsVacuum Service Option Installation Model 3300 PA Installation ChecklistPercent Oxygen Analyzer Operation Operation IntroductionOperation Model 3300PA Using the Function and Data Entry ButtonsSetting the Analysis Ranges Set Alarm Setting the Alarm SetpointsPercent Oxygen Analyzer Operation HI Range LO RangeSensor Fail Alarm Calibration Selecting a Fixed Range or AutorangingSupplementary Information Percent Oxygen Analyzer Maintenance MaintenanceRemove Power to Unit before replacing the fuse Replacing the FuseMaintenance Model 3300PA Sensor Installation or ReplacementWhen to Replace a Sensor Ordering and Handling of Spare SensorsRemoving the Micro-Fuel Cell Installing a Micro-Fuel CellCell Warranty Conditions Percent Oxygen Analyzer Appendix Appendix SpecificationsAppendix Model 3300PA Spare Parts ListMiscellaneous Reference DrawingAppendix Model 3300PA