Caractéristiques
Remarque : le masque ne contient pas de latex, de PVC ni de DEHP.
Courbe Pression/Débit
ventilation |
| 70.0 |
|
|
|
|
|
|
|
|
|
|
|
|
|
| Pression | Débit |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||
| 60.0 |
|
|
|
|
|
|
|
|
|
|
|
|
|
| (cm H2O) | (± 6,0 l/min) | |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||
|
| 50.0 |
|
|
|
|
|
|
|
|
|
|
|
|
|
| 4 | 22 |
orificesde | (l/min) | 40.0 |
|
|
|
|
|
|
|
|
|
|
|
|
|
| 8 | 32 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| 16 | 48 | ||
|
| 30.0 |
|
|
|
|
|
|
|
|
|
|
|
|
|
| 12 | 41 |
aux |
| 20.0 |
|
|
|
|
|
|
|
|
|
|
|
|
|
| 20 | 54 |
| 10.0 |
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||||
Débit |
| 0.0 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| 2 | 4 | 6 | 8 | 10 | 12 | 14 | 16 | 18 | 20 | 22 | 24 | 26 | 28 | 30 |
|
Pression au masque (cm H2O)
Remarque : en raison des variations de fabrication, le débit aux orifices de ventilation peut différer des valeurs indiquées sur le schéma.
Information concernant l’espace mort
L’espace mort correspond au volume vide entre le masque et le coude. L'espace mort maximal du Mirage Liberty pour toutes les combinaisons de taille est 151 ml.
Normes
Marquage CE conformément à la directive CE 93/42/CEE, Classe IIa. ISO 14971, ISO
Pression de traitement
4 à 20 cm H2O
Résistance
Chute de pression mesurée (nominale)
à50 l/min : 0,5 cm H2O
à100 l/min : 1,6 cm H2O
Niveau acoustique
Le niveau acoustique pondéré A du masque est 28 dBA.
Conditions ambiantes
Température de fonctionnement : +5°C à +40°C
Humidité de fonctionnement : 10 à 95 % d'humidité relative (sans condensation)
Température de stockage et de transport :
Humidité de stockage et de transport : 10 à 95 % d'humidité relative (sans condensation)
Français
31