Omron HBP-1300 instruction manual System Settings, Menu Mode, Press the Mode button

Page 18

System Settings

The system settings are divided into two modes, “Menu Mode” and “Utility Mode”.

Menu Mode

“Menu Mode” allows you to configure the “Initial Inflation Pressure Value” and “Auscultation Mode” settings.

Auscultation Mode

Initial Inflation

 

Pressure Value

1. Press the [MODE] button.

The “Initial Inflation Pressure Value” setting screen appears.

2.Press the [ ] or [ ] button to change the setting.

Press the [ ] button to change the setting in the order “280”, “250”, “220”, “AUTO”, “280”...

Press the [ ] button to change the setting in the order “220”, “250”, “280”, “AUTO”, “220”...

When the “Initial Inflation Pressure Value” is set, inflation takes place at a fixed speed to the set value and thus is quicker.

“AUTO” estimates the systolic blood pressure during inflation and automatically inflates the cuff to a suitable value.

When using “220”, “250”, or “280” mmHg, select the value that is 30 to 40 mmHg higher than the estimated systolic blood pressure.

3. When you have completed the “Initial Inflation Pressure Value” setting, press the [MODE] button.

The “Auscultation Mode” settings screen appears.

“ON” or “OFF” appears.

4. Press the [ ] or [ ] button to set to “ON” or “OFF”.

When set to “ON”, SYS and DIA can be recorded using auscultation measurement. For information on auscultation measurement, see page 22.

5. When you have completed the “Auscultation Mode” setting, press the [MODE] button.

“0” appears.

To start blood pressure measurement, press the [START/STOP] button.

Note:

The body movement detection function is disabled while “Auscultation Mode” is in use.

If the power is turned off, the settings revert to the factory settings.

“Initial Inflation Pressure Value” changes to “AUTO”, and “Auscultation Mode” changes to “OFF”.

16

Image 18
Contents HBP-1300 Contents Introduction Intended UseExemptions General Information Setup Usage warnings and cautionsBefore use / during use Page Cleaning Maintenance and inspectionRechargeable battery Non-Invasive Blood Pressure Nibp measurement For cleaning, see Using the Unit Components of the ProductOptions Features of the Product Optional Medical AccessoriesKey features, benefits, look Features / Functions of Unit Meaning of the SymbolsFront and back of unit LCD Display DIALow battery Installing the Battery PackBattery life Auto Power OffBattery level Connecting the AC AdapterBattery charging AC powerPress the Mode button System SettingsMenu Mode Press the or button to change the settingPress the or button to change the auto power off Setting Utility ModeConfirm that the device is switched off Check the Pressure AccuracyNon-Invasive Blood Pressure Nibp Measurement Non-Invasive Pressure Measurement PrinciplesSelecting the cuff Connecting the cuffCuff Selection and Connection Cuff name Arm circumference InchTo 2 cm Align the artery mark Index Artery with the brachial arteryMeasurement results are displayed Taking the Measurement in Auscultation Mode Irregular pulse wave detection functionBody movement detection function Stopping the Measurement Displaying Last ReadingCleaning of the Device MaintenanceMaintenance Inspection and Safety Management Accessory CareAfter turning on the power Before turning on the powerWhen turning on the power Check before UseTroubleshooting Abnormal measurement value Measurement value is questionableList of Error Codes Error Priority Description Points to checkMedium Internal hardware error Contact Omron Healthcare Disposal Other problemsError Priority Cause Solution Code DescriptionTechnical Specifications HBP-1300 SpecificationsFactory Default Settings Power supplyEnvironmental Conditions Non-Invasive Blood Pressure NibpPage Manufacturer’s Declaration Emission Test Compliance Electromagnetic EnvironmentRecommend separation distance Separation distance according to frequency of transmitter Production facility ManufacturerEU-representative Subsidiary Omron Medizintechnik Handelsgesellschaft mbH