Recommended Separation Distances:

Recommended separation distance between portable and mobile RF communications equipment and the HBP-1300

The HBP-1300 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the HBP-1300 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the HBP-1300 as recommended below, according to the maximum output power of the communications equipment.

 

Separation distance according to frequency of transmitter

Rated maximum output

m

 

 

 

 

 

 

power of transmitter

 

 

 

 

 

 

 

 

 

150 kHz to 80 MHz

80 MHz to 800 MHz

800 MHz to 2.5 GHz

(W)

 

 

 

 

 

 

 

 

 

 

d = 1.2

P

 

d = 1.2

P

 

d = 2.3

P

 

 

 

 

 

 

 

 

 

 

 

0.01

0.12

 

 

0.12

 

 

0.23

 

 

 

 

 

 

 

 

 

 

 

 

0.1

0.38

 

 

0.38

 

 

0.73

 

 

 

 

 

 

 

 

 

 

 

 

1

1.2

 

 

1.2

 

 

2.3

 

 

 

 

 

 

 

 

 

 

 

 

10

3.8

 

 

3.8

 

 

7.3

 

 

 

 

 

 

 

 

 

 

 

 

100

12

 

 

12

 

 

23

 

 

 

 

 

 

 

 

 

 

 

 

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be determined using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.

Note1: At 80MHz and 800MHz, the separation distance for the higher frequency range applies

Note2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

EN

35

Image 37
Contents HBP-1300 Contents Intended Use IntroductionExemptions General Information Usage warnings and cautions SetupBefore use / during use Page Maintenance and inspection CleaningRechargeable battery Non-Invasive Blood Pressure Nibp measurement For cleaning, see Components of the Product Using the UnitOptions Optional Medical Accessories Features of the ProductKey features, benefits, look Meaning of the Symbols Features / Functions of UnitFront and back of unit DIA LCD DisplayBattery life Installing the Battery PackLow battery Auto Power OffBattery charging Connecting the AC AdapterBattery level AC powerMenu Mode System SettingsPress the Mode button Press the or button to change the settingConfirm that the device is switched off Utility ModePress the or button to change the auto power off Setting Check the Pressure AccuracyNon-Invasive Pressure Measurement Principles Non-Invasive Blood Pressure Nibp MeasurementCuff Selection and Connection Connecting the cuffSelecting the cuff Cuff name Arm circumference InchAlign the artery mark Index Artery with the brachial artery To 2 cmMeasurement results are displayed Irregular pulse wave detection function Taking the Measurement in Auscultation ModeBody movement detection function Displaying Last Reading Stopping the MeasurementMaintenance Inspection and Safety Management MaintenanceCleaning of the Device Accessory CareWhen turning on the power Before turning on the powerAfter turning on the power Check before UseTroubleshooting Measurement value is questionable Abnormal measurement valueError Priority Description Points to check List of Error CodesMedium Internal hardware error Contact Omron Healthcare Error Priority Cause Solution Code Other problemsDisposal DescriptionFactory Default Settings SpecificationsTechnical Specifications HBP-1300 Power supplyNon-Invasive Blood Pressure Nibp Environmental ConditionsPage Emission Test Compliance Electromagnetic Environment Manufacturer’s DeclarationRecommend separation distance Separation distance according to frequency of transmitter EU-representative ManufacturerProduction facility Subsidiary Omron Medizintechnik Handelsgesellschaft mbH