Omron HBP-1300 Maintenance Inspection and Safety Management, Cleaning of the Device

Page 26

Maintenance

Maintenance Inspection and Safety Management

The HBP-1300 must be maintained to ensure functionality and to secure the safety of patients and operators. Daily checks and maintenance should be performed by the operator. (page 25)

In addition, qualified personnel are necessary to maintain the performance and the safety, and to conduct periodic inspections. We recommend that the verification test be performed at least once a year.

Example of connection for pressure accuracy confirmation:

1.Display the pressure accuracy confirmation screen as explained in “Utility Mode” on page 17.

Display “0” in the pressure accuracy confirmation screen.

Display the pressure value.

2.Connect the blood pressure monitor, the calibrated reference pressure gauge, and the cuff and inflation bulb.

3.Check the pressure value of the blood pressure monitor and the pressure value of the calibrated reference pressure gauge.

Note:

Make sure that the blood pressure monitor reading is within ±3 mmHg compared to the calibrated reference pressure gauge.

To rapidly deflate the cuff, press the [] button.

To repeat “Pressure Accuracy Confirmation”, turn off the power and repeat the procedure from step 1 in “Utility Mode” on page 17.

Cleaning of the Device

Cleaning and disinfecting should be performed in accordance with your facility’s infection control practice.

Surface cleaning

Wipe with a cloth that has been moistened with isopropyl alcohol diluted to 50 v/v%, or ethyl alcohol (disinfection alcohol) diluted to 80 v/v% or less and wrung out.

Do not wipe the DC jack or allow it to become wet.

Removing dust

Use a moistened cotton bud to remove dust that has accumulated on the vent ports.

Service

The device requires no routine service other than cleaning, and visually checking the cuffs, tubing, etc.

Caution

Do not sterilize by autoclave or gas sterilization (EOG, formaldehyde gas, high-concentration ozone, etc.).

If using an antiseptic solution for cleaning, follow the instructions of the manufacturer.

Accessory Care

Non-Invasive Blood Pressure Measurement (NIBP)

Cuff

Wipe clean on the surface of the cuff with a cloth moistened with a 70 v/v% dilution of isopropyl alcohol, or a 80 v/v% or less dilution of disinfection ethanol (ethyl alcohol).

Do not allow any liquids inside the cuff. If a liquid gets in the cuff, dry the inside well.

24

Image 26
Contents HBP-1300 Contents Exemptions IntroductionIntended Use General Information Setup Usage warnings and cautionsBefore use / during use Page Cleaning Maintenance and inspectionRechargeable battery Non-Invasive Blood Pressure Nibp measurement For cleaning, see Options Using the UnitComponents of the Product Key features, benefits, look Features of the ProductOptional Medical Accessories Front and back of unit Features / Functions of UnitMeaning of the Symbols LCD Display DIALow battery Installing the Battery PackBattery life Auto Power OffBattery level Connecting the AC AdapterBattery charging AC powerPress the Mode button System SettingsMenu Mode Press the or button to change the settingPress the or button to change the auto power off Setting Utility ModeConfirm that the device is switched off Check the Pressure AccuracyNon-Invasive Blood Pressure Nibp Measurement Non-Invasive Pressure Measurement PrinciplesSelecting the cuff Connecting the cuffCuff Selection and Connection Cuff name Arm circumference InchTo 2 cm Align the artery mark Index Artery with the brachial arteryMeasurement results are displayed Body movement detection function Taking the Measurement in Auscultation ModeIrregular pulse wave detection function Stopping the Measurement Displaying Last ReadingCleaning of the Device MaintenanceMaintenance Inspection and Safety Management Accessory CareAfter turning on the power Before turning on the powerWhen turning on the power Check before UseTroubleshooting Abnormal measurement value Measurement value is questionableMedium Internal hardware error Contact Omron Healthcare List of Error CodesError Priority Description Points to check Disposal Other problemsError Priority Cause Solution Code DescriptionTechnical Specifications HBP-1300 SpecificationsFactory Default Settings Power supplyEnvironmental Conditions Non-Invasive Blood Pressure NibpPage Manufacturer’s Declaration Emission Test Compliance Electromagnetic EnvironmentRecommend separation distance Separation distance according to frequency of transmitter Production facility ManufacturerEU-representative Subsidiary Omron Medizintechnik Handelsgesellschaft mbH