Omron HBP-1300 Check before Use, Before turning on the power, When turning on the power

Page 27

Check before Use

Before conducting safety checks, be sure to implement the items in the “Cleaning of the Device” and “Accessory Care” sections. (page 24)

Before turning on the power

Before turning on the power, check for the following

External appearance

￿The device or accessories are not deformed due to falling or other impact.

￿The device is not dirty.

￿The device is not wet.

AC adapter

￿The AC adapter is firmly connected to the connector on the device.

￿There are no heavy objects lying on the AC adapter cable.

￿The AC adapter cable is not damaged (core-wire exposure, breaks, etc.).

When turning on the power

When turning on the power, check the following

Display/lamp

￿When the [START/STOP] button is pressed to turn on the power, the screen below appears and the alarm lamp lights (page 13).

The alarm lamp is lit up.

EN

After turning on the power

After turning on the power, check for the following

External appearance

￿There is no smoke or odor coming from the device.

￿The device is not making any unusual noises.

Buttons

￿Press each button and check that it works.

Non-invasive blood pressure (NIBP)

￿Make sure that a suitable OMRON GS CUFF is attached (one that fits the circumference of the patient’s arm).

￿The cuff tube is firmly connected.

￿The person checking the cuff should wrap the cuff around arm, perform cuff measurement and check to see that blood pressure is in the vicinity of normal measurements.

￿While measurement is in progress, bend the relevant arm and move body to halt discharge and during this halt check that cuff pressure does not drop.

25

Image 27
Contents HBP-1300 Contents Introduction Intended UseExemptions General Information Usage warnings and cautions SetupBefore use / during use Page Maintenance and inspection CleaningRechargeable battery Non-Invasive Blood Pressure Nibp measurement For cleaning, see Using the Unit Components of the ProductOptions Features of the Product Optional Medical AccessoriesKey features, benefits, look Features / Functions of Unit Meaning of the SymbolsFront and back of unit DIA LCD DisplayAuto Power Off Installing the Battery PackBattery life Low batteryAC power Connecting the AC AdapterBattery charging Battery levelPress the or button to change the setting System SettingsMenu Mode Press the Mode buttonCheck the Pressure Accuracy Utility ModeConfirm that the device is switched off Press the or button to change the auto power off SettingNon-Invasive Pressure Measurement Principles Non-Invasive Blood Pressure Nibp MeasurementCuff name Arm circumference Inch Connecting the cuffCuff Selection and Connection Selecting the cuffAlign the artery mark Index Artery with the brachial artery To 2 cmMeasurement results are displayed Taking the Measurement in Auscultation Mode Irregular pulse wave detection functionBody movement detection function Displaying Last Reading Stopping the MeasurementAccessory Care MaintenanceMaintenance Inspection and Safety Management Cleaning of the DeviceCheck before Use Before turning on the powerWhen turning on the power After turning on the powerTroubleshooting Measurement value is questionable Abnormal measurement valueList of Error Codes Error Priority Description Points to checkMedium Internal hardware error Contact Omron Healthcare Description Other problemsError Priority Cause Solution Code DisposalPower supply SpecificationsFactory Default Settings Technical Specifications HBP-1300Non-Invasive Blood Pressure Nibp Environmental ConditionsPage Emission Test Compliance Electromagnetic Environment Manufacturer’s DeclarationRecommend separation distance Separation distance according to frequency of transmitter Subsidiary Omron Medizintechnik Handelsgesellschaft mbH ManufacturerEU-representative Production facility