Non-Invasive Blood Pressure (NIBP) measurement

Warning

If a cuff is used on a patient with an infection, treat the cuff as medical waste, or disinfect before reuse.

If frequently performing NIBP measurement using a cuff over an extended period of time, periodically check the patient’s circulation. In addition, wrap the cuff as indicated in the cautionary points in this manual.

Do not connect the NIBP cuff or cuff joint to a luer lock adapter.

Do not bend cuff tube during inflation and deflation, particularly after a change of body position.

Do not wrap the cuff on the following parts:

-An upper arm on which intravenous drip or a blood transfusion is being performed.

-An upper arm on which an SpO2 sensor, IBP catheter, or other instrument is attached.

-An upper arm with a shunt for hemodialysis

If measuring blood pressure with the cuff wrapped on the arm on the side of the body where a mastectomy was performed, check the patient’s condition.

Caution

NIBP measurement should be performed on the upper arm.

During NIBP, stop excessive body movement by the patient and minimize trembling.

If a doctor has indicated that the patient has hemorrhagic diathesis or hypercoagulability, check the condition of the arm after measurement.

Use the appropriate cuff size to ensure correct measurements. If too large a cuff is used, the measured blood pressure value tends to be lower than the actual blood pressure. If too small a cuff is used, the measured blood pressure value tends to be higher.

Before and during measurement, verify that none of the following apply to the patient:

-The part where the cuff is wrapped is at a different height than the heart.

(A difference of 10 cm (4 inches) in height may cause a variation in the blood pressure value of up to 7 or 8 mmHg.)

-Body movement or conversing during measurement.

-Cuff wrapped over thick clothing.

-Pressure on the arm due to a rolled up sleeve.

In the case of a cuff for adults, the cuff should be wrapped to a tightness that allows two fingers to be inserted in between the cuff and the arm.

The accuracy of a flashing measurement value that is out of the measurement range cannot be guaranteed. Always check the patient’s condition before deciding what steps to take.

Do not use the cuff if it is damaged or has holes.

Only OMRON GS CUFF can be used with this device.

8

Image 10
Contents HBP-1300 Contents Intended Use IntroductionExemptions General Information Setup Usage warnings and cautionsBefore use / during use Page Cleaning Maintenance and inspectionRechargeable battery Non-Invasive Blood Pressure Nibp measurement For cleaning, see Components of the Product Using the UnitOptions Optional Medical Accessories Features of the ProductKey features, benefits, look Meaning of the Symbols Features / Functions of UnitFront and back of unit LCD Display DIALow battery Installing the Battery PackBattery life Auto Power OffBattery level Connecting the AC AdapterBattery charging AC powerPress the Mode button System SettingsMenu Mode Press the or button to change the settingPress the or button to change the auto power off Setting Utility ModeConfirm that the device is switched off Check the Pressure AccuracyNon-Invasive Blood Pressure Nibp Measurement Non-Invasive Pressure Measurement PrinciplesSelecting the cuff Connecting the cuffCuff Selection and Connection Cuff name Arm circumference InchTo 2 cm Align the artery mark Index Artery with the brachial arteryMeasurement results are displayed Irregular pulse wave detection function Taking the Measurement in Auscultation ModeBody movement detection function Stopping the Measurement Displaying Last ReadingCleaning of the Device MaintenanceMaintenance Inspection and Safety Management Accessory CareAfter turning on the power Before turning on the powerWhen turning on the power Check before UseTroubleshooting Abnormal measurement value Measurement value is questionableError Priority Description Points to check List of Error CodesMedium Internal hardware error Contact Omron Healthcare Disposal Other problemsError Priority Cause Solution Code DescriptionTechnical Specifications HBP-1300 SpecificationsFactory Default Settings Power supplyEnvironmental Conditions Non-Invasive Blood Pressure NibpPage Manufacturer’s Declaration Emission Test Compliance Electromagnetic EnvironmentRecommend separation distance Separation distance according to frequency of transmitter Production facility ManufacturerEU-representative Subsidiary Omron Medizintechnik Handelsgesellschaft mbH