Omron HBP-1300 Connecting the AC Adapter, Battery charging, Battery level, AC power

Page 17

Battery charging

State

LCD contents and operations

Icon

Charging

The icon flashes.

 

 

 

 

Fully charged

The icon is lit.

 

(Charging completed)

 

 

 

 

 

 

Problem with the

Error message is displayed.

-

battery

 

 

 

 

 

Battery level

Battery level

LCD contents and operations

Icon

Fully charged

The icon is lit.

 

The unit can be used.

 

 

 

 

 

 

Battery level is 20%.

The icon flashes (E40 error not displayed).

 

The unit can be used.

 

 

 

 

 

 

 

The icon flashes (E40 error displayed).

 

 

The unit cannot be used.

-

Battery level is 5%.

 

If the unit is being continuously used, the power will be

 

 

 

 

automatically turned off in 30 seconds.

 

 

 

 

Connecting the AC Adapter

AC power

Verify that the power outlet supplies the specified voltage and frequency (100 - 240 V AC, 50/60 Hz).

EN

Connect the AC adapter to the DC jack on the unit and the power outlet.

Note:

When the battery pack is installed

If there is no problem with the following.

-AC adapter

-DC jack

-Power outlet

-Battery

And the icon does not flash, contact OMRON HEALTHCARE.

15

Image 17
Contents HBP-1300 Contents Exemptions IntroductionIntended Use General Information Usage warnings and cautions SetupBefore use / during use Page Maintenance and inspection CleaningRechargeable battery Non-Invasive Blood Pressure Nibp measurement For cleaning, see Options Using the UnitComponents of the Product Key features, benefits, look Features of the ProductOptional Medical Accessories Front and back of unit Features / Functions of Unit Meaning of the Symbols DIA LCD DisplayBattery life Installing the Battery PackLow battery Auto Power OffBattery charging Connecting the AC AdapterBattery level AC powerMenu Mode System SettingsPress the Mode button Press the or button to change the settingConfirm that the device is switched off Utility ModePress the or button to change the auto power off Setting Check the Pressure AccuracyNon-Invasive Pressure Measurement Principles Non-Invasive Blood Pressure Nibp MeasurementCuff Selection and Connection Connecting the cuffSelecting the cuff Cuff name Arm circumference InchAlign the artery mark Index Artery with the brachial artery To 2 cmMeasurement results are displayed Body movement detection function Taking the Measurement in Auscultation ModeIrregular pulse wave detection function Displaying Last Reading Stopping the MeasurementMaintenance Inspection and Safety Management MaintenanceCleaning of the Device Accessory CareWhen turning on the power Before turning on the powerAfter turning on the power Check before UseTroubleshooting Measurement value is questionable Abnormal measurement valueMedium Internal hardware error Contact Omron Healthcare List of Error CodesError Priority Description Points to check Error Priority Cause Solution Code Other problemsDisposal DescriptionFactory Default Settings SpecificationsTechnical Specifications HBP-1300 Power supplyNon-Invasive Blood Pressure Nibp Environmental ConditionsPage Emission Test Compliance Electromagnetic Environment Manufacturer’s DeclarationRecommend separation distance Separation distance according to frequency of transmitter EU-representative ManufacturerProduction facility Subsidiary Omron Medizintechnik Handelsgesellschaft mbH