Omron HBP-1300 instruction manual Stopping the Measurement, Displaying Last Reading

Page 25

Stopping the Measurement

To stop measurement while measurement is in progress, press [START/STOP] button.

Displaying Last Reading

The previous measurement value (systolic blood pressure, diastolic blood pressure, and pulse rate) and whether or not an irregular pulse wave was detected can be displayed. In the case of auscultation measurement, the auscultation icon appears.

Press the [ ] or [ ] button to display the previous data. This function is also available when the device is switched off.

If an irregular pulse wave or body movement was detected, the irregular pulse wave icon appears.

When the data was measured in “Auscultation Mode”, the auscultation icon appears.

Lights while the previous data are being displayed.

If held down at least 3 seconds when a measurement result is displayed, the measured data is cleared without being stored.

Note:

When the monitor is left idle for one minute, the backlight will disappear.

EN

23

Image 25
Contents HBP-1300 Contents Intended Use IntroductionExemptions General Information Usage warnings and cautions SetupBefore use / during use Page Maintenance and inspection CleaningRechargeable battery Non-Invasive Blood Pressure Nibp measurement For cleaning, see Components of the Product Using the UnitOptions Optional Medical Accessories Features of the ProductKey features, benefits, look Meaning of the Symbols Features / Functions of UnitFront and back of unit DIA LCD DisplayBattery life Installing the Battery PackLow battery Auto Power OffBattery charging Connecting the AC AdapterBattery level AC powerMenu Mode System SettingsPress the Mode button Press the or button to change the settingConfirm that the device is switched off Utility ModePress the or button to change the auto power off Setting Check the Pressure AccuracyNon-Invasive Pressure Measurement Principles Non-Invasive Blood Pressure Nibp MeasurementCuff Selection and Connection Connecting the cuffSelecting the cuff Cuff name Arm circumference Inch Align the artery mark Index Artery with the brachial artery To 2 cmMeasurement results are displayed Irregular pulse wave detection function Taking the Measurement in Auscultation ModeBody movement detection function Displaying Last Reading Stopping the MeasurementMaintenance Inspection and Safety Management MaintenanceCleaning of the Device Accessory CareWhen turning on the power Before turning on the powerAfter turning on the power Check before UseTroubleshooting Measurement value is questionable Abnormal measurement valueError Priority Description Points to check List of Error CodesMedium Internal hardware error Contact Omron Healthcare Error Priority Cause Solution Code Other problemsDisposal DescriptionFactory Default Settings SpecificationsTechnical Specifications HBP-1300 Power supplyNon-Invasive Blood Pressure Nibp Environmental ConditionsPage Emission Test Compliance Electromagnetic Environment Manufacturer’s DeclarationRecommend separation distance Separation distance according to frequency of transmitter EU-representative ManufacturerProduction facility Subsidiary Omron Medizintechnik Handelsgesellschaft mbH