Page 24

Irregular pulse wave detection function

If the pulse wave interval becomes irregular during measurement, the irregular pulse wave detection icon will light to notify you.

Body movement detection function

If body movement is detected during measurement, deflation stops for 5 seconds. The irregular pulse wave icon appears in the measurement result display.

￿Deflation stopped

After 5 seconds, measurement resumes, and an attempt is made to complete measurement in one cycle.

Taking the Measurement in “Auscultation Mode”

Use a stethoscope to perform measurement.

When set to “ON”, SYS and DIA can be recorded using auscultation measurement. To determine SYS and DIA during measurement, press the [MODE] button.

SYS is registered the first time and DIA is registered the second time you press the [MODE] button during deflation.

After DIA is determined, the cuff rapidly deflates and SYS and DIA are displayed as the measurement results.

The pulse rate does not appear in the measurement result display.

During deflation, the [ ] button can be held down to re-inflate, or the [ ] button can be held down to deflate faster.

Stored data that was measured in “Auscultation Mode” is displayed when the auscultation icon is lit.

￿During re-inflation

Note:

The body movement detection function is disabled while “Auscultation Mode” is in use. For Auscultation Mode settings, see page 16.

22

Image 24
Contents HBP-1300 Contents Introduction Intended UseExemptions General Information Setup Usage warnings and cautionsBefore use / during use Page Cleaning Maintenance and inspectionRechargeable battery Non-Invasive Blood Pressure Nibp measurement For cleaning, see Using the Unit Components of the ProductOptions Features of the Product Optional Medical AccessoriesKey features, benefits, look Features / Functions of Unit Meaning of the SymbolsFront and back of unit LCD Display DIAInstalling the Battery Pack Battery lifeLow battery Auto Power OffConnecting the AC Adapter Battery chargingBattery level AC powerSystem Settings Menu ModePress the Mode button Press the or button to change the settingUtility Mode Confirm that the device is switched offPress the or button to change the auto power off Setting Check the Pressure AccuracyNon-Invasive Blood Pressure Nibp Measurement Non-Invasive Pressure Measurement Principles Connecting the cuff Cuff Selection and Connection Selecting the cuff Cuff name Arm circumference InchTo 2 cm Align the artery mark Index Artery with the brachial arteryMeasurement results are displayed Taking the Measurement in Auscultation Mode Irregular pulse wave detection functionBody movement detection function Stopping the Measurement Displaying Last ReadingMaintenance Maintenance Inspection and Safety ManagementCleaning of the Device Accessory CareBefore turning on the power When turning on the powerAfter turning on the power Check before UseTroubleshooting Abnormal measurement value Measurement value is questionableList of Error Codes Error Priority Description Points to checkMedium Internal hardware error Contact Omron Healthcare Other problems Error Priority Cause Solution CodeDisposal DescriptionSpecifications Factory Default SettingsTechnical Specifications HBP-1300 Power supplyEnvironmental Conditions Non-Invasive Blood Pressure NibpPage Manufacturer’s Declaration Emission Test Compliance Electromagnetic EnvironmentRecommend separation distance Separation distance according to frequency of transmitter Manufacturer EU-representativeProduction facility Subsidiary Omron Medizintechnik Handelsgesellschaft mbH