Environmental Conditions

Operational temperature and

Temperature range: 10 to 40°C (50 to 104°F)

humidity

Humidity range: 30 to 85%RH (not condensed)

 

Atmospheric pressure: 700 to 1060hPa

 

 

Storage and transportation

Temperature range: -20 to 60°C (-4 to 140°F)

 

Humidity range: 10 to 95%RH (not condensed)

 

Atmospheric pressure: 500 to 1060hPa

 

 

Non-Invasive Blood Pressure (NIBP)

Measurement technology

Oscillometric

Measurement method

Dynamic Linear Deflation method

 

 

Pressure display range

0 to 300 mmHg

 

 

Pressure display accuracy

Within ±3 mmHg

 

 

NIBP measurement range

SYS 60 to 250 mmHg

 

DIA 40 to 200 mmHg

 

Pulse rate 40 to 200 /min

 

 

NIBP accuracy*

Maximum mean error within ±5mmHg

 

Maximum standard deviation within 8mmHg

 

 

Pulse rate accuracy

Within ±5 % of reading

 

 

Reference standard:

EN1060-1:1995+A2:2009

 

EN1060-3:1997+A2:2009

 

ISO81060-1:2007

 

 

*Comparison with auscultation method performed by a trained professional. DIA determined by the auscultation method is “K5”.

Note: Specifications may be changed without prior notice.

EN

This blood pressure monitor fulfils the requirements of the EC directive 93/42/EEC (Medical Device Directive). It also conforms to the European standard EN 1060, Non-invasive Sphygmomanometers Part 1: General Requirements and Part 3: Additional Requirements for Electromechanical Blood Pressure Measuring Systems.

31

Image 33
Contents HBP-1300 Contents Introduction Intended UseExemptions General Information Usage warnings and cautions SetupBefore use / during use Page Maintenance and inspection CleaningRechargeable battery Non-Invasive Blood Pressure Nibp measurement For cleaning, see Using the Unit Components of the ProductOptions Features of the Product Optional Medical AccessoriesKey features, benefits, look Features / Functions of Unit Meaning of the SymbolsFront and back of unit DIA LCD DisplayBattery life Installing the Battery PackLow battery Auto Power OffBattery charging Connecting the AC AdapterBattery level AC powerMenu Mode System SettingsPress the Mode button Press the or button to change the settingConfirm that the device is switched off Utility ModePress the or button to change the auto power off Setting Check the Pressure AccuracyNon-Invasive Pressure Measurement Principles Non-Invasive Blood Pressure Nibp MeasurementCuff Selection and Connection Connecting the cuffSelecting the cuff Cuff name Arm circumference InchAlign the artery mark Index Artery with the brachial artery To 2 cmMeasurement results are displayed Taking the Measurement in Auscultation Mode Irregular pulse wave detection functionBody movement detection function Displaying Last Reading Stopping the MeasurementMaintenance Inspection and Safety Management MaintenanceCleaning of the Device Accessory CareWhen turning on the power Before turning on the powerAfter turning on the power Check before UseTroubleshooting Measurement value is questionable Abnormal measurement valueList of Error Codes Error Priority Description Points to checkMedium Internal hardware error Contact Omron Healthcare Error Priority Cause Solution Code Other problemsDisposal DescriptionFactory Default Settings SpecificationsTechnical Specifications HBP-1300 Power supplyNon-Invasive Blood Pressure Nibp Environmental ConditionsPage Emission Test Compliance Electromagnetic Environment Manufacturer’s DeclarationRecommend separation distance Separation distance according to frequency of transmitter EU-representative ManufacturerProduction facility Subsidiary Omron Medizintechnik Handelsgesellschaft mbH