Omron HBP-1300 Disposal, Other problems, Description, Error Priority Cause Solution Code

Page 31

Other problems

Error

Priority

Cause

Solution

code

 

 

 

 

 

Inflated the cuff to 300 mmHg or more

When inflating in “Auscultation Mode”,

 

 

release the button when the pressure

 

 

during inflation in “Auscultation Mode”.

 

 

reaches the desired value.

 

 

 

 

 

 

 

E3

Low

 

If this occurs during measurement, repeat

 

measurement.

 

 

 

 

 

Over-inflation occurs

If this occurs when not performing

 

 

 

measurement, contact OMRON

 

 

 

HEALTHCARE.

 

 

 

 

E40

Medium

The battery is depleted.

Recharge the battery, or replace with a new

battery. (page 14)

 

 

 

 

 

 

 

E41

Medium

Battery failed to charge.

Try charging again. If continues to fail,

replace with a new battery. (page 14)

 

 

 

 

 

 

 

 

 

 

Replace the battery with a new battery. If the

E42

Medium

Battery voltage error

error continues, contact OMRON

 

 

 

HEALTHCARE.

 

 

 

 

Disposal

Description

As there is a risk of environmental pollution, follow your applicable national and local legal regulations regarding disposal or recycling of this equipment and batteries.

The main constituents of each part are listed in the table below. As there is a risk of infection, do not recycle patient attachments such as cuffs, but dispose of them as instructed by your facility’s procedures and applicable regulations.

Item

Parts

Material

 

 

Box

Cardboard

EN

Package

Cushion

Cardboard

 

Bag

PE

 

Main unit and

Enclosure

ABS, PC , SR

 

accessories

Internal parts

General electronic components

 

 

Outer tube

PVC

 

Battery pack

Cell batteries

Nickel-metal hydride

 

 

Internal parts

General electronic components

 

29

Image 31
Contents HBP-1300 Contents Intended Use IntroductionExemptions General Information Usage warnings and cautions SetupBefore use / during use Page Maintenance and inspection CleaningRechargeable battery Non-Invasive Blood Pressure Nibp measurement For cleaning, see Components of the Product Using the UnitOptions Optional Medical Accessories Features of the ProductKey features, benefits, look Meaning of the Symbols Features / Functions of UnitFront and back of unit DIA LCD DisplayAuto Power Off Installing the Battery PackBattery life Low batteryAC power Connecting the AC AdapterBattery charging Battery levelPress the or button to change the setting System SettingsMenu Mode Press the Mode buttonCheck the Pressure Accuracy Utility ModeConfirm that the device is switched off Press the or button to change the auto power off SettingNon-Invasive Pressure Measurement Principles Non-Invasive Blood Pressure Nibp MeasurementCuff name Arm circumference Inch Connecting the cuffCuff Selection and Connection Selecting the cuffAlign the artery mark Index Artery with the brachial artery To 2 cmMeasurement results are displayed Irregular pulse wave detection function Taking the Measurement in Auscultation ModeBody movement detection function Displaying Last Reading Stopping the MeasurementAccessory Care MaintenanceMaintenance Inspection and Safety Management Cleaning of the DeviceCheck before Use Before turning on the powerWhen turning on the power After turning on the powerTroubleshooting Measurement value is questionable Abnormal measurement valueError Priority Description Points to check List of Error CodesMedium Internal hardware error Contact Omron Healthcare Description Other problemsError Priority Cause Solution Code DisposalPower supply SpecificationsFactory Default Settings Technical Specifications HBP-1300Non-Invasive Blood Pressure Nibp Environmental ConditionsPage Emission Test Compliance Electromagnetic Environment Manufacturer’s DeclarationRecommend separation distance Separation distance according to frequency of transmitter Subsidiary Omron Medizintechnik Handelsgesellschaft mbH ManufacturerEU-representative Production facility