Utility Mode

“Utility Mode” allows you to configure the “Auto Power Off” and “Pressure Accuracy Confirmation” setting.

1. Confirm that the device is switched off.

If the power is on, hold down the [START/STOP] button for at least 3 seconds to turn off the power.

2. Hold down the [MODE] button until the “Auto Power Off” setting screen appears.

The “Initial Inflation Pressure Value” setting screen appears, and changes to the “Auto Power Off” setting screen appears.

3. Press the [ ] or [ ] button to change the auto power off

“5” or “10”

setting.

appears.

The “5 min.” or “10 min.” setting is entered.

When the unit is powered by the battery pack and is not used for the set time (“5 min.” or “10 min.”), the power automatically turns off to save battery power.

If a mid-priority alarm other than a low battery error (E40 error) has occurred, the power does not turn off automatically.

4.When you have completed the “Auto Power Off” setting, press the [MODE] button.

The “Pressure Accuracy Confirmation” screen appears. “0” appears.

“0” appears.

EN

5. Check the Pressure Accuracy.

Add pressure externally as explained on page 24.

Compare with the displayed value and verify that there is no problem.

6. When you have completed “Pressure Accuracy Confirmation”, press the [START/STOP] button.

The power turns off.

17

Image 19
Contents HBP-1300 Contents Intended Use IntroductionExemptions General Information Usage warnings and cautions SetupBefore use / during use Page Maintenance and inspection CleaningRechargeable battery Non-Invasive Blood Pressure Nibp measurement For cleaning, see Components of the Product Using the UnitOptions Optional Medical Accessories Features of the ProductKey features, benefits, look Meaning of the Symbols Features / Functions of UnitFront and back of unit DIA LCD DisplayAuto Power Off Installing the Battery PackBattery life Low batteryAC power Connecting the AC AdapterBattery charging Battery levelPress the or button to change the setting System SettingsMenu Mode Press the Mode buttonCheck the Pressure Accuracy Utility ModeConfirm that the device is switched off Press the or button to change the auto power off SettingNon-Invasive Pressure Measurement Principles Non-Invasive Blood Pressure Nibp MeasurementCuff name Arm circumference Inch Connecting the cuffCuff Selection and Connection Selecting the cuffAlign the artery mark Index Artery with the brachial artery To 2 cmMeasurement results are displayed Irregular pulse wave detection function Taking the Measurement in Auscultation ModeBody movement detection function Displaying Last Reading Stopping the MeasurementAccessory Care MaintenanceMaintenance Inspection and Safety Management Cleaning of the DeviceCheck before Use Before turning on the powerWhen turning on the power After turning on the powerTroubleshooting Measurement value is questionable Abnormal measurement valueError Priority Description Points to check List of Error CodesMedium Internal hardware error Contact Omron Healthcare Description Other problemsError Priority Cause Solution Code DisposalPower supply SpecificationsFactory Default Settings Technical Specifications HBP-1300Non-Invasive Blood Pressure Nibp Environmental ConditionsPage Emission Test Compliance Electromagnetic Environment Manufacturer’s DeclarationRecommend separation distance Separation distance according to frequency of transmitter Subsidiary Omron Medizintechnik Handelsgesellschaft mbH ManufacturerEU-representative Production facility