Omron HBP-1300 To 2 cm, Align the artery mark Index Artery with the brachial artery

Page 22

Applying the Cuff to the Patient

1. Wrap on a bare arm or over thin clothing.

Wrap the cuff on a bare arm or over thin clothing. Thick clothing or a rolled up sleeve will cause inaccurate blood pressure measurements.

The device can be used on either the right or left arm.

2. Align the artery mark “INDEXARTERY” with the brachial artery.

Run the cuff tube out the peripheral side with no bends (the brachial artery is on the inner side of the patient’s upper arm).

Make sure that “INDEXARTERY” is within the “RANGE” and the lower edge of the cuff is 1 to 2 cm from the inner side of the elbow joint.

If “INDEXARTERY” is outside the “RANGE”, error in the blood pressure value will increase. In this case, use a different cuff size.

As a guideline for the tightness of the cuff, It should be possible to insert about two fingers under the cuff.

1.

2.

3.

1 to 2 cm

1 to 2 cm

3.During measurement, keep the brachial artery on which the cuff is wrapped at the same height as the right atrium of the heart.

20

Image 22
Contents HBP-1300 Contents Intended Use IntroductionExemptions General Information Setup Usage warnings and cautionsBefore use / during use Page Cleaning Maintenance and inspectionRechargeable battery Non-Invasive Blood Pressure Nibp measurement For cleaning, see Components of the Product Using the UnitOptions Optional Medical Accessories Features of the ProductKey features, benefits, look Meaning of the Symbols Features / Functions of UnitFront and back of unit LCD Display DIALow battery Installing the Battery PackBattery life Auto Power OffBattery level Connecting the AC AdapterBattery charging AC powerPress the Mode button System SettingsMenu Mode Press the or button to change the settingPress the or button to change the auto power off Setting Utility ModeConfirm that the device is switched off Check the Pressure AccuracyNon-Invasive Blood Pressure Nibp Measurement Non-Invasive Pressure Measurement PrinciplesSelecting the cuff Connecting the cuffCuff Selection and Connection Cuff name Arm circumference InchTo 2 cm Align the artery mark Index Artery with the brachial arteryMeasurement results are displayed Irregular pulse wave detection function Taking the Measurement in Auscultation ModeBody movement detection function Stopping the Measurement Displaying Last ReadingCleaning of the Device MaintenanceMaintenance Inspection and Safety Management Accessory CareAfter turning on the power Before turning on the powerWhen turning on the power Check before UseTroubleshooting Abnormal measurement value Measurement value is questionableError Priority Description Points to check List of Error CodesMedium Internal hardware error Contact Omron Healthcare Disposal Other problemsError Priority Cause Solution Code DescriptionTechnical Specifications HBP-1300 SpecificationsFactory Default Settings Power supplyEnvironmental Conditions Non-Invasive Blood Pressure NibpPage Manufacturer’s Declaration Emission Test Compliance Electromagnetic EnvironmentRecommend separation distance Separation distance according to frequency of transmitter Production facility ManufacturerEU-representative Subsidiary Omron Medizintechnik Handelsgesellschaft mbH