Omron HBP-1300 instruction manual Measurement results are displayed

Page 23

Caution

Make sure the cuff is wrapped in the correct arm position and is at the same height as the heart.

A difference of 10 cm (4 inches) in height may cause a variation in the blood pressure value of up to 7 - 8 mmHg.

Note:

If measurement is difficult due to arrhythmia, use a different blood pressure measurement method.

If the patient has acute inflammation, a pyogenic ailment, or an external wound at the location where the cuff is to be wrapped, follow the instructions of the doctor.

Non-Invasive Blood Pressure (NIBP) measurement is performed by compressing the upper arm. Some people may experience intense pain, or transient spotting caused by subcutaneous hemorrhaging may appear. The spotting will disappear over time, however, if it is possible that this will disturb the patient, try the following technique:

-Wrap a thin towel or cloth (one layer) under the cuff.

If the towel or cloth is too thick, there will be insufficient cuff compression and the blood pressure value will measure high.

If the patient moves or the cuff is touched, this may be falsely detected as a pulse and over-inflation will occur.

Do not inflate the cuff when it is not wrapped on the upper arm. This may damage the cuff.

Taking the Measurement in “Manual Mode”

1. Press the [START/STOP] button.

Blood pressure measurement is performed once.

2. The measurement results are displayed.

If a measurement value is outside the corresponding range below, the value will flash. SYS: 59 mmHg or less, or 251 mmHg or higher.

DIA: 39 mmHg or less, or 201 mmHg or higher.

PULSE: 39 bpm or less, or 201 bpm or higher.

￿ Normal measurement

￿ Measurement error /

 

failure

EN

Caution

The accuracy of a flashing measurement value that is outside the measurement range is not guaranteed. Always check the patient’s condition before deciding what steps to take.

Note:

If inflation is insufficient, inflation may restart automatically while measurement is in progress.

21

Image 23
Contents HBP-1300 Contents Exemptions IntroductionIntended Use General Information Usage warnings and cautions SetupBefore use / during use Page Maintenance and inspection CleaningRechargeable battery Non-Invasive Blood Pressure Nibp measurement For cleaning, see Options Using the UnitComponents of the Product Key features, benefits, look Features of the ProductOptional Medical Accessories Front and back of unit Features / Functions of UnitMeaning of the Symbols DIA LCD DisplayAuto Power Off Installing the Battery PackBattery life Low batteryAC power Connecting the AC AdapterBattery charging Battery levelPress the or button to change the setting System SettingsMenu Mode Press the Mode buttonCheck the Pressure Accuracy Utility ModeConfirm that the device is switched off Press the or button to change the auto power off SettingNon-Invasive Pressure Measurement Principles Non-Invasive Blood Pressure Nibp MeasurementCuff name Arm circumference Inch Connecting the cuffCuff Selection and Connection Selecting the cuffAlign the artery mark Index Artery with the brachial artery To 2 cmMeasurement results are displayed Body movement detection function Taking the Measurement in Auscultation ModeIrregular pulse wave detection function Displaying Last Reading Stopping the MeasurementAccessory Care MaintenanceMaintenance Inspection and Safety Management Cleaning of the DeviceCheck before Use Before turning on the powerWhen turning on the power After turning on the powerTroubleshooting Measurement value is questionable Abnormal measurement valueMedium Internal hardware error Contact Omron Healthcare List of Error CodesError Priority Description Points to check Description Other problemsError Priority Cause Solution Code DisposalPower supply SpecificationsFactory Default Settings Technical Specifications HBP-1300Non-Invasive Blood Pressure Nibp Environmental ConditionsPage Emission Test Compliance Electromagnetic Environment Manufacturer’s DeclarationRecommend separation distance Separation distance according to frequency of transmitter Subsidiary Omron Medizintechnik Handelsgesellschaft mbH ManufacturerEU-representative Production facility