A&D UA-1020CN instruction manual EMC Table, Cispr

Page 28

EMC TABLE

Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided in the following. Portable and mobile RF communication equipment (e.g. cell phones) can affect Medical Electrical Equipment. The use of accessories and cables other than those specified may result in increased emissions or decreased immunity of the unit.

Guidance and manufacturer’s declaration – electromagnetic emissions

The A&D unit is intended for use in the electromagnetic environment specified below. The customer or the user of the A&D unit should assure that it is used in such an environment.

Emissions test

Compliance

Electromagnetic environment – guidance

RF emissions

Group 1

The A&D unit uses RF energy only for its

CISPR 11

 

internal function. Therefore, its RF emissions

 

 

are very low and are not likely to cause any

 

 

interference in nearby electronic equipment.

 

 

 

RF emissions

Class B

The A&D unit is suitable for use in all

CISPR 11

 

establishments,

including

domestic

Harmonic emissions

Class A

establishments and those directly connected

IEC 61000-3-2

 

to the public low-voltage power supply

Voltage fluctuations /

Complies

network that supplies buildings used for

flicker emissions

 

domestic purposes.

 

 

IEC 61000-3-3

 

 

 

 

Recommended separation distances between portable and mobile RF communications equipment and the A&D unit

The A&D unit is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the A&D unit can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the A&D unit as recommended below, according to the maximum output power of the communications equipment.

Rated maximum

Separation distance according to frequency of transmitter

output

 

m

 

 

 

power of transmitter

150 kHz to 80 MHz

80 MHz

to

800

800 MHz to 2.5 GHz

W

d = 1.2 P

MHz

 

 

d = 2.3 P

d = 1.2

P

 

 

 

 

 

 

 

 

0.01

0.12

0.12

 

0.23

0.1

0.38

0.38

 

0.73

1

1.2

1.2

 

 

2.3

10

3.8

3.8

 

 

7.3

100

12

12

 

 

23

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where p is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.

NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

E-27

A&D Medical (Toll-Free): 1-888-726-9966

Image 28
Contents Models UA-1020 + UA-1020CN This Monitor is Clinically Validated PrecautionsPreliminary Remarks Important Information Medical Designed for Life Table of Contents What Display Symbols Mean Symbols that are printed on the device caseWhat Display Symbols Mean SYSMonitor Components DisplayHOW UA-1020 Works Before YOU StartSelecting the Correct Cuff Size Adjusting the Clock Adjusting the PressureTricheck Mode Tips for Blood Pressure Monitoring Taking Your Blood PressureNOW YOU are Ready Follow These Simple Steps Press Start Measurement Using Tricheck Press the Start buttonMeasurement with SET Pressure Measurement with Desired Systolic PressureWhat is AN Irregular Heartbeat About Cuff Inflation MeterAbout the Pressure Rating Indicator About Memory Press the , or Start button to turn the device offConnecting the AC Adapter to the Monitor Using the Monitor with AC AdapterDisconnecting the AC Adapter from the Monitor What is Blood PRESSURE? About Blood PressureWhat Affects Blood PRESSURE? What Causes Variations in Blood PRESSURE?100 Assessing High Blood Pressure for AdultsWHY Measure Blood Pressure AT HOME? What is HYPERTENSION?Can Hypertension be CONTROLLED? Keys to Successful Monitoring HOW do I Record MY Blood PRESSURE? Establishing Baseline MeasurementsCustomer Support Tools Online Troubleshooting Problem Probable Cause Corrective ActionSpecifications 0366 U.S.AEMC Table CisprRecommended separation distance Should Contact Information CanadaDate SYS/DIA Pulse Pouls Fecha SIS/DIA Pulso Canada Auto Control Medical Toll-Free Modèles UA-1020 + UA-1020CN CE Tensiomètre EST Approuvé Cliniquement Remarques PréliminairesMesures DE Sécurité ÉLECTRIQUE, Respecter Attentivement CES DirectivesRenseignements Importants Canada Auto Control Medical sans frais Table DES Matières Signification DES Symboles Symboles imprimés sur le boîtier de lappareilSYS Écran Pièces DU TensiomètreFonctionnement DE Lappareil UA-1020 Avant DE CommencerSélection DU Brassard DE Taille Appropriée Réglage DE Lhorloge Réglage DE LA TensionMode Tricheck Le mode TriCheck Il est désactivé par défautConseils Pour LA Prise DE LA Tension Artérielle Prendre LA Tension ArtérielleVous Êtes Maintenant Prêt Procéder Comme Suit Appuyer sur Start Mesure À Laide DE Tricheck RemarquesMesure Avec Tension Définie Mesure Avec Tension Systolique DésiréeQU’EST-CE QU’UNE Arythmie ? Indicateur DE Gonflage DU BrassardHypertension Niveau Préhypertension Normale Exemple AU Sujet DE LA Mémoire Mesures. En cas dabsence de données, « -- » safficheConnexion DE L’ADAPTATEUR CA AU Tensiomètre Utilisation DU Tensiomètre Avec UN Adaptateur CADéconnexion DE L’ADAPTATEUR CA DU Tensiomètre Renseignements Relatifs À LA Tension Artérielle QU’EST-CE QUE LA Tension Artérielle ?Renseignements Relatifs À LA Tension Artérielle Évaluation DE Lhypertension Chez L’ADULTEDéfinition DE Lhypertension EST-IL Possible DE Contrôler Lhypertension ?Renseignements Relatifs À LA Tension Artérielle Comment DOIS-JE Inscrire MA Tension Artérielle ? Établissement DE Mesures DE RéférenceOutils Dassistance Clientèle EN Ligne Dépannage Problème Cause probable Mesure correctiveCaractéristiques Techniques Tension 0 à 299 mm de HgTableau DE Compatibilité Électromagnétique Test d’émissions ConformitéDistance déloignement recommandée Test d’immunité Niveau de test Coordonnées Tous les autres paysCanada Auto Control Medical Toll-Free Medical Toll-Free Modelos UA-1020 + UA-1020CN Este Monitor Está Clínicamente Validado Observaciones PreliminaresPrecauciones Importantes Instrucciones DE Seguridad Guarde EstasInformación Importante PrecaucionesRápido. Fácil. Preciso Índice Significado DE LOS Símbolos DE LA Pantalla Símbolos impresos en el estuche del dispositivoCuando el símbolo parpadeé Pantalla Componentes DEL MonitorCómo Funciona UA-1020 Antes DE ComenzarSelección DEL Brazalete DE Tamaño Correcto Ajustes DEL Reloj Ajustes DE LA PresiónModo Tricheck Consejos Para EL Control DE LA Presión Arterial Medición DE LA Presión ArterialAhora Está Listo Siga Estos Simples Pasos Presione Start Medición CON Tricheck Presione el botón StartMedición CON Presión DE Referencia Medición CON Presión Sistólica DeseadaQUÉ ES UN Latido Irregular DEL Corazón Acerca DEL Medidor DE Ingreso DE Aire EN EL BrazaleteAcerca DEL Indicador DE Clasificacion DE Presión Acerca DE LA Memoria Para Conectar EL Adaptador DE CA AL Monitor USO DEL Monitor CON Adaptador DE CAPara Desconectar EL Adaptador DE CA DEL Monitor Acerca DE LA Presión Arterial ¿QUÉ ES LA Presión ARTERIAL?¿QUÉ Afecta a LA Presión ARTERIAL? ¿QUÉ Factores Provocan Variaciones EN LA Presión ARTERIAL?Evaluación DE LA Hipertensión Arterial EN Adultos Igual a¿ES Posible Mantener LA Hipertensión Bajo CONTROL? ¿QUÉ ES LA HIPERTENSIÓN?¿POR QUÉ Tomarse LA Presión Arterial EN CASA? Claves Para UN Control Satisfactorio ¿CÓMO Llevo UN Registro DE MI Presión ARTERIAL? Cómo Establecer Mediciones DE ReferenciaAsistencia AL Cliente EN Línea Resolución DE Problemas Problema Causa probable Medida correctivaEspecificaciones Presión diastólica 40 200 mmHgTabla EMC = 1,2 = 2,3 100Guía y declaración del fabricante inmunidad electromagnética Línea a línea Información DE Contacto En los Estados Unidos de AméricaDate SYS/DIA Pulse Pouls Fecha SIS/DIA Pulso Canada Auto Control Medical Toll-Free Auto Control Medical, Inc