Replacing the Air Filter
To replace the air filter:
1. Remove the air filter cover at the back of the S8 Elite II.
Air filter cover
2.Remove and discard the old air filter.
3.Insert a new filter with the blue tinted side facing out from the device.
4.Replace the air filter cover.
Inspect the air filter every month to check if it is blocked by dirt or contains holes. With normal use of an S8 Elite II, the air filter needs to be replaced every six months (or more often if your device is in a dusty environment).
WARNING
! • Do not wash the air filter. The air filter is not washable or reusable.
• The air filter cover protects the device in the event of accidental liquid spillage onto the device. Ensure the air filter and air filter cover are fitted at all times.
Servicing
CAUTION
! • Do not attempt to open the S8 Elite II case. There are no user serviceable parts inside.
• Inspection and repair should only be performed by an authorised agent. Under no circumstances should you attempt to service or repair the flow generator yourself.
This product (S8 Elite II) should be inspected by an authorised ResMed service centre 5 years from the date of manufacture. Prior to this, the device is intended to provide safe and reliable operation provided that it is operated and maintained in accordance with the instructions provided by ResMed. Applicable ResMed warranty details are provided with the device at the time of original supply. Of course, as with all electrical devices, if any irregularity becomes apparent, you should exercise caution and have the device inspected by an authorised ResMed service centre.
If you feel that your flow generator is not performing properly, see “Troubleshooting” on page 27.
Cleaning and Maintenance | 24 |