Problem | Possible Cause |
Display error | You have experienced |
message: High | excessively high leak levels |
leak in last | during the night. |
session |
|
Solution
Check that your air tubing is connected properly and that your mask does not leak excessively.
Use the
If this message appears again, contact your clinician.
Display error | You have experienced |
message: HIGH | excessively high leak levels for |
LEAK!!! | more than 20 seconds. |
Adjust Mask |
|
Check that your air tubing is connected properly and that your mask does not leak excessively.
Use the
If this message appears again, contact your clinician.
Display error | Component failure. | Return your S8 Elite II for |
message: |
| servicing. |
Exxxx (where xxxx |
|
|
defines an error) |
|
|
|
|
|
The following message is displayed on the LCD after you try to update settings using the Data Card:
Card Error
Remove Card
The Data Card is not inserted correctly.
The Data Card is only partly inserted.
Ensure that the Data Card is inserted with the arrow facing up and pointing towards the device. Ensure that the device is switched on and in standby mode before you insert the Data Card.
Ensure that the Data Card is inserted into the slot on the Data Card module as far as it will go.
You may have removed the Data Card before settings were copied to the S8 Elite II.
The card is not a ResScan Data Card.
Reinsert the Data Card and wait for the
Settings Success
Remove Card
message to appear on the LCD.
Only use a ResScan Data Card.
Troubleshooting 29