ResMed 338425/1 manual Leak

Page 35

Problem

Possible Cause

Display error

You have experienced

message: High

excessively high leak levels

leak in last

during the night.

session

 

Solution

Check that your air tubing is connected properly and that your mask does not leak excessively.

Use the mask-fitting feature to help you to fit your mask properly.

If this message appears again, contact your clinician.

Display error

You have experienced

message: HIGH

excessively high leak levels for

LEAK!!!

more than 20 seconds.

Adjust Mask

 

Check that your air tubing is connected properly and that your mask does not leak excessively.

Use the mask-fitting feature to help you to fit your mask properly.

If this message appears again, contact your clinician.

Display error

Component failure.

Return your S8 Elite II for

message:

 

servicing.

Exxxx (where xxxx

 

 

defines an error)

 

 

 

 

 

The following message is displayed on the LCD after you try to update settings using the Data Card:

Card Error

Remove Card

The Data Card is not inserted correctly.

The Data Card is only partly inserted.

Ensure that the Data Card is inserted with the arrow facing up and pointing towards the device. Ensure that the device is switched on and in standby mode before you insert the Data Card.

Ensure that the Data Card is inserted into the slot on the Data Card module as far as it will go.

You may have removed the Data Card before settings were copied to the S8 Elite II.

The card is not a ResScan Data Card.

Reinsert the Data Card and wait for the

Settings Success

Remove Card

message to appear on the LCD.

Only use a ResScan Data Card.

Troubleshooting 29

Image 35
Contents S8 Elite II System Page S8 Elite II System Blank Module Contents Replacing the Air Filter Servicing Your Responsibility IntroductionContraindications Medical InformationAbout the S8 Elite Page Adverse Effects Masks S8 Elite II SystemS8 Elite ResScan Data Card Module AccessoriesHumidifier Assemble the Mask How to Use the S8 EliteSet Up the S8 Elite SmartStart Start TreatmentStopping Treatment Excelnt Page Storing the HumidAire How to Attach a HumidifierAttach a HumidAire 3i Humidifier Keypad Keys How to use the KeypadLCD Screen How to Change the Ramp Time How to Change the SettingsHow to Select the Mask Type How to Use the Standard MenuHow to Change the Expiratory Pressure Relief Setting Other viewable information Shows the detailed menu How to Use the Detailed MenuHow to Change the SmartStart Setting Settings MenuHow to Change the Tube Length How to Add a Humidifier or Front CapMessages on the S8 Elite II LCD Results MenuOptions Menu Servicing MenuReplace S8 Modules Using a Blank ModuleUsing a ResScan Data Card Module Using the Data Card to Update Settings Page Periodic Cleaning Cleaning and MaintenanceDaily Cleaning Weekly CleaningServicing Replacing the Air FilterFrequently Asked Questions Can I use my S8 Elite II if mains AC power is not available? No display TroubleshootingSystem ERR nnnn Leak Settings Success Remove Card S8 Elite II no Accessories System SpecificationsData Card Module All Devices where applicableData Card Symbols which appear on the device R001-307/1 System Specifications Portable and mobile RF communications Separation distance according to frequency of transmitter Limited Warranty Index PCB