Page 43

Recommended separation distances between portable and mobile RF communications equipment and the S8 series of devices

The S8 series of devices is intended for use in an environment in which radiated RF disturbances are controlled. The customer or the user of the S8 device can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the S8 device as recommended below, according to the maximum output power of the communications equipment.

 

Separation distance according to frequency of transmitter

 

m

 

 

 

 

 

 

Rated maximum output

150 kHz to 80 MHz

80 MHz to 800 MHz

800 MHz to 2.5 GHz

power of transmitter

d = 1.17 P

d = 0.35 P

d = 0.7 P

W

 

 

 

 

0.01

0.17

0.04

0.07

 

 

 

 

0.1

0.37

0.11

0.22

 

 

 

 

1

1.17

0.35

0.7

 

 

 

 

10

3.69

1.11

2.21

 

 

 

 

100

11.70

3.50

7.0

 

 

 

 

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be determined using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.

NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

System Specifications

37

Image 43
Contents S8 Elite II System Page S8 Elite II System Blank Module Contents Replacing the Air Filter Servicing Your Responsibility IntroductionAbout the S8 Elite Medical InformationContraindications Page Adverse Effects S8 Elite S8 Elite II SystemMasks Humidifier AccessoriesResScan Data Card Module Set Up the S8 Elite How to Use the S8 EliteAssemble the Mask Stopping Treatment Start TreatmentSmartStart Excelnt Page Attach a HumidAire 3i Humidifier How to Attach a HumidifierStoring the HumidAire LCD Screen How to use the KeypadKeypad Keys How to Change the Ramp Time How to Change the SettingsHow to Change the Expiratory Pressure Relief Setting How to Use the Standard MenuHow to Select the Mask Type Other viewable information Shows the detailed menu How to Use the Detailed MenuHow to Change the SmartStart Setting Settings MenuHow to Change the Tube Length How to Add a Humidifier or Front CapMessages on the S8 Elite II LCD Results MenuOptions Menu Servicing MenuReplace Using a ResScan Data Card Module Using a Blank ModuleS8 Modules Using the Data Card to Update Settings Page Periodic Cleaning Cleaning and MaintenanceDaily Cleaning Weekly CleaningServicing Replacing the Air FilterFrequently Asked Questions Can I use my S8 Elite II if mains AC power is not available? No display TroubleshootingSystem ERR nnnn Leak Settings Success Remove Card S8 Elite II no Accessories System SpecificationsData Card All Devices where applicableData Card Module Symbols which appear on the device R001-307/1 System Specifications Portable and mobile RF communications Separation distance according to frequency of transmitter Limited Warranty Index PCB