Page 8

Medical Information

About the S8 Elite II

The S8 Elite II continuous positive airway pressure (CPAP) system is indicated for the treatment of obstructive sleep apnoea (OSA) in adult patients.

The S8 Elite II CPAP system is intended for home and hospital use.

Contraindications

The S8 Elite II is not a life support device and may stop operating during power failure or certain fault conditions. It should not be used by patients who are dependent on continuous therapy.

Positive airway pressure therapy may be contraindicated in some patients with the following pre-existing conditions:

severe bullous lung disease

pneumothorax

pathologically low blood pressure

dehydration

cerebrospinal fluid leak, recent cranial surgery or trauma.

Warnings

The entire manual should be read before using the S8 Elite II.

Advice contained in this manual should not supersede instructions given by the prescribing physician.

A patient should not connect a device to the data communication port unless instructed to do so by their health care provider or physician. Only ResMed products are designed to be connected to the data communication port. Connecting other devices could result in injury, or damage to the S8 Elite II device.

The S8 Elite II should be used with masks (and connectors)1 recommended by ResMed, or by a physician or respiratory therapist. A mask should not be used unless the S8 Elite II device is turned on and operating properly. The vent hole or holes associated with the mask should never be blocked.

Explanation: The S8 Elite II is intended to be used with special masks (or connectors)1 which have vent holes to allow continuous flow of air out of the

mask.

When the device is turned on and functioning properly, new air from the device flushes the exhaled air out through the mask vent holes. However, when the device is not operating, insufficient fresh air will be provided

1Ports may be incorporated into the mask or in connectors that are near the mask.

Medical Information

2

Image 8
Contents S8 Elite II System Page S8 Elite II System Blank Module Contents Replacing the Air Filter Servicing Introduction Your ResponsibilityContraindications Medical InformationAbout the S8 Elite Page Adverse Effects Masks S8 Elite II SystemS8 Elite ResScan Data Card Module AccessoriesHumidifier Assemble the Mask How to Use the S8 EliteSet Up the S8 Elite SmartStart Start TreatmentStopping Treatment Excelnt Page Storing the HumidAire How to Attach a HumidifierAttach a HumidAire 3i Humidifier Keypad Keys How to use the KeypadLCD Screen How to Change the Settings How to Change the Ramp TimeHow to Select the Mask Type How to Use the Standard MenuHow to Change the Expiratory Pressure Relief Setting Other viewable information How to Use the Detailed Menu Shows the detailed menuSettings Menu How to Change the Tube LengthHow to Add a Humidifier or Front Cap How to Change the SmartStart SettingResults Menu Options MenuServicing Menu Messages on the S8 Elite II LCDReplace S8 Modules Using a Blank ModuleUsing a ResScan Data Card Module Using the Data Card to Update Settings Page Cleaning and Maintenance Daily CleaningWeekly Cleaning Periodic CleaningReplacing the Air Filter ServicingFrequently Asked Questions Can I use my S8 Elite II if mains AC power is not available? Troubleshooting No displaySystem ERR nnnn Leak Settings Success Remove Card System Specifications S8 Elite II no AccessoriesData Card Module All Devices where applicableData Card Symbols which appear on the device R001-307/1 System Specifications Portable and mobile RF communications Separation distance according to frequency of transmitter Limited Warranty Index PCB