through the mask, and the exhaled air may be rebreathed. Rebreathing of exhaled air for longer than several minutes can, in some circumstances, lead to suffocation. This applies to most models of CPAP devices.

In the event of power failure or machine malfunction, remove the mask.

The S8 Elite II can be set to deliver pressures up to 20 cm H2O. In the unlikely event of certain fault conditions, pressures up to 30 cm H2O are possible.

If oxygen is used with this device, the oxygen flow must be turned off when the device is not operating. If oxygen has been left on while the device is not operating, disconnect the oxygen and wait 30 minutes before turning on the device.

Explanation: When the CPAP device is not in operation and the oxygen flow is left on, oxygen delivered into the air delivery tubing may accumulate within the CPAP machine enclosure and create a risk of fire. This applies to most types of CPAP machines.

Oxygen supports combustion. Oxygen should not be used while you are smoking or in the presence of an open flame.

Always ensure airflow is being generated by the device before the oxygen supply is turned on.

Always turn the oxygen supply off before stopping the airflow from the device.

Note: At a fixed rate of supplemental oxygen flow, the inhaled oxygen concentration will vary, depending on where the oxygen is introduced, the pressure settings, patient breathing pattern, mask selection and leak rate.

Do not use the S8 Elite II if there are obvious external defects or unexplained changes in performance.

Do not open the S8 Elite II case. There are no user serviceable parts inside. Repairs and internal servicing should only be performed by an authorised service agent.

Explosion hazard—do not use in the vicinity of flammable anesthetics.

Cautions

At low pressures, the flow through the exhalation ports of your mask may not clear all exhaled gas from the tubing. Some rebreathing may occur.

The airflow for breathing produced by this device can be as much as 6°C higher than the temperature of the room. Caution should be exercised if the room temperature is warmer than 32°C.

When AC mains power (100–240V AC) is not available, always use a ResMed DC-12 converter. (The DC-12 converter is available as an optional accessory.)

If supplied, the ResScan™ Data Card module is designed for use with specific ResMed devices.

Only use the ResScan Data Card module (if supplied) as specified in this manual.

Medical Information

3

Image 9
Contents S8 Elite II System Page S8 Elite II System Blank Module Contents Replacing the Air Filter Servicing Your Responsibility IntroductionMedical Information About the S8 EliteContraindications Page Adverse Effects S8 Elite II System S8 EliteMasks Accessories HumidifierResScan Data Card Module How to Use the S8 Elite Set Up the S8 EliteAssemble the Mask Start Treatment Stopping TreatmentSmartStart Excelnt Page How to Attach a Humidifier Attach a HumidAire 3i HumidifierStoring the HumidAire How to use the Keypad LCD ScreenKeypad Keys How to Change the Ramp Time How to Change the SettingsHow to Use the Standard Menu How to Change the Expiratory Pressure Relief SettingHow to Select the Mask Type Other viewable information Shows the detailed menu How to Use the Detailed MenuHow to Change the Tube Length Settings MenuHow to Add a Humidifier or Front Cap How to Change the SmartStart SettingOptions Menu Results MenuServicing Menu Messages on the S8 Elite II LCDReplace Using a Blank Module Using a ResScan Data Card ModuleS8 Modules Using the Data Card to Update Settings Page Daily Cleaning Cleaning and MaintenanceWeekly Cleaning Periodic CleaningServicing Replacing the Air FilterFrequently Asked Questions Can I use my S8 Elite II if mains AC power is not available? No display TroubleshootingSystem ERR nnnn Leak Settings Success Remove Card S8 Elite II no Accessories System SpecificationsAll Devices where applicable Data CardData Card Module Symbols which appear on the device R001-307/1 System Specifications Portable and mobile RF communications Separation distance according to frequency of transmitter Limited Warranty Index PCB