Guidance and manufacturer’s declaration—electromagnetic immunity

The S8 series of devices is intended for use in the electromagnetic environment specified below. The customer or the user of the S8 device should assure that the device is used in such an environment.

Immunity

IEC60601-1-2

Compliance

Electromagnetic environment—

test

test level

level

guidance

 

 

 

 

 

 

 

Portable and mobile RF communications

 

 

 

equipment should be used no closer to any

 

 

 

part of the S8 device, including cables, than

 

 

 

the recommended separation distance

 

 

 

calculated from the equation applicable to the

 

 

 

frequency of the transmitter.

 

 

 

Recommended separation distance

Conducted RF

3 Vrms

3 Vrms

d = 1.17 P

IEC 61000-4-6

150 kHz to 80 MHz

 

 

Radiated RF

10 V/m

10 V/m

d = 0.35 P 80 MHz to 800 MHz

IEC 61000-4-3

80 MHz to 2.5 GHz

 

 

 

 

 

d = 0.70 P 800 MHz to 2.5 GHz

 

 

 

where P is the maximum output power rating

 

 

 

of the transmitter in watts (W) according to the

 

 

 

transmitter manufacturer and d is the

 

 

 

recommended separation distance in meters

 

 

 

(m).

 

 

 

Field strengths from fixed RF transmitters, as

 

 

 

determined by an electromagnetic site

 

 

 

survey,a should be less than the compliance

 

 

 

level in each frequency range.b

 

 

 

Interference may occur in the vicinity of

 

 

 

equipment marked with the following symbol:

 

 

 

 

NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

aField strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the S8 device is used exceeds the applicable RF compliance level above, the S8 device should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the S8 device.

bOver the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.

System Specifications

36

Image 42
Contents S8 Elite II System Page S8 Elite II System Blank Module Contents Replacing the Air Filter Servicing Introduction Your ResponsibilityMedical Information About the S8 EliteContraindications Page Adverse Effects S8 Elite II System S8 EliteMasks Accessories HumidifierResScan Data Card Module How to Use the S8 Elite Set Up the S8 EliteAssemble the Mask Start Treatment Stopping TreatmentSmartStart Excelnt Page How to Attach a Humidifier Attach a HumidAire 3i HumidifierStoring the HumidAire How to use the Keypad LCD ScreenKeypad Keys How to Change the Settings How to Change the Ramp TimeHow to Use the Standard Menu How to Change the Expiratory Pressure Relief SettingHow to Select the Mask Type Other viewable information How to Use the Detailed Menu Shows the detailed menuHow to Add a Humidifier or Front Cap Settings MenuHow to Change the Tube Length How to Change the SmartStart SettingServicing Menu Results MenuOptions Menu Messages on the S8 Elite II LCDReplace Using a Blank Module Using a ResScan Data Card ModuleS8 Modules Using the Data Card to Update Settings Page Weekly Cleaning Cleaning and MaintenanceDaily Cleaning Periodic CleaningReplacing the Air Filter ServicingFrequently Asked Questions Can I use my S8 Elite II if mains AC power is not available? Troubleshooting No displaySystem ERR nnnn Leak Settings Success Remove Card System Specifications S8 Elite II no AccessoriesAll Devices where applicable Data CardData Card Module Symbols which appear on the device R001-307/1 System Specifications Portable and mobile RF communications Separation distance according to frequency of transmitter Limited Warranty Index PCB