ResMed 338425/1 manual Troubleshooting, No display

Page 33

Troubleshooting

If there is a problem, try the following suggestions. If the problem cannot be solved, contact your equipment supplier or ResMed. Do not attempt to open the device.

Problem

Possible Cause

Solution

 

 

 

No display.

Power not connected.

Ensure the appropriate power

 

 

cord is connected to your

 

 

S8 Elite II and the power switch

 

 

(if available) is on.

 

 

 

Insufficient air

Ramp time is in use.

Wait for air pressure to build up

delivered from

 

or change ramp time.

S8 Elite II.

Air filter is dirty.

Replace air filter.

 

 

Air tubing is kinked or

Straighten or replace tubing.

 

punctured.

 

 

Air tubing not connected

Check air tubing.

 

properly.

 

 

Mask and headgear not

Adjust position of mask and

 

positioned correctly.

headgear.

 

Cap(s) missing from access

Replace cap(s).

 

port(s) on mask.

 

 

Pressure required for treatment

See your clinician to adjust the

 

may have changed.

pressure.

 

 

 

S8 Elite II does not

Power cord not connected

Connect power cord firmly at

start when you

properly.

both ends.

breathe into the

Power outlet may be faulty.

Try another power outlet.

mask.

 

 

 

SmartStart not on.

Enable SmartStart.

 

Leak Alert has been enabled;

Disable Leak Alert to enable

 

SmartStart has automatically

SmartStart.

 

been disabled.

 

 

Use of a Mirage or Ultra Mirage

SmartStart does not work with a

 

Full Face Mask.

full face mask as the anti-

 

 

asphyxia valve will not allow

 

 

sufficiently high pressure on

 

 

exhalation.

Breath is not deep enough to trigger SmartStart.

Take a deep breath in and out through the mask.

Troubleshooting 27

Image 33
Contents S8 Elite II System Page S8 Elite II System Blank Module Contents Replacing the Air Filter Servicing Your Responsibility IntroductionMedical Information About the S8 EliteContraindications Page Adverse Effects S8 Elite II System S8 EliteMasks Accessories HumidifierResScan Data Card Module How to Use the S8 Elite Set Up the S8 EliteAssemble the Mask Start Treatment Stopping TreatmentSmartStart Excelnt Page How to Attach a Humidifier Attach a HumidAire 3i HumidifierStoring the HumidAire How to use the Keypad LCD ScreenKeypad Keys How to Change the Ramp Time How to Change the SettingsHow to Use the Standard Menu How to Change the Expiratory Pressure Relief SettingHow to Select the Mask Type Other viewable information Shows the detailed menu How to Use the Detailed MenuHow to Change the Tube Length Settings MenuHow to Add a Humidifier or Front Cap How to Change the SmartStart SettingOptions Menu Results MenuServicing Menu Messages on the S8 Elite II LCDReplace Using a Blank Module Using a ResScan Data Card ModuleS8 Modules Using the Data Card to Update Settings Page Daily Cleaning Cleaning and MaintenanceWeekly Cleaning Periodic CleaningServicing Replacing the Air FilterFrequently Asked Questions Can I use my S8 Elite II if mains AC power is not available? No display TroubleshootingSystem ERR nnnn Leak Settings Success Remove Card S8 Elite II no Accessories System SpecificationsAll Devices where applicable Data CardData Card Module Symbols which appear on the device R001-307/1 System Specifications Portable and mobile RF communications Separation distance according to frequency of transmitter Limited Warranty Index PCB